- [ ] Pairing
- [ ] Pairing works with `combo-scripter-v2` branch
- [ ] Bolusing
- [ ] Cancelling bolus at various stages and checking error message (if any) and check
no bolus or a partial bolus was delivered
- [ ] Cancelling a bolus 0.2 U before its finished (cancelling requires pressing up/down
3s, so the cancellation will be too late, but the code should handle it and report
the full bolus being delivered
- [ ] Low cartridge alarm during bolus
- [ ] must be confirmed by AAPS
- [ ] bolus must have been fully delivered by pump
- [ ] bolus must have been added to DB
- [ ] the confirmed pump warning must be raised as a notification in AAPS
- [ ] notification must also appear on smartwatch
- [ ] An error during bolus must yield an error in AAPS
- [ ] An error during bolus must yield a notification on a smartwatch
- [ ] Bolusing e.g. 4 U if reservoir has only 2 U left must be rejected
- [ ] Pressing a button on the pump before bolus delivery started must be handled gracefully
- [ ] Same as above, but moving pump out of range
- [ ] Pressing a button on the pump after bolus delivery has started must be handled
gracefully (bolus finished and AAPS must correctly add it the the DB)
- [ ] Same as above, but moving pump out of range
- [ ] Test the highest bolus you'd ever give yourself (AAPS has a configurable limit and the pump
has a limit which can be configured with the Config SW)
- [ ] Test a bolus exceeding the pumps max bolus; should show a generic error message
- [ ] BT disconnect issues
- [ ] Moving pump out of reach when setting TBR causes "TBR cancelled" alarm on pump.
When putting pump close to phone, AAPS must confirm the alert and successfully
retry setting TBR (reconnects are a best-effort kind of a thing, so this might not always work)
- [ ] Same as above while bolusing must report an error and NOT retry the command
- [ ] AAPS start
- [ ] Starting AAPS without a reachable pump must show something sensible in the Combo tab
(that is not hang indefinitely with "initializing" activity)
- [ ] Starting AAPS without a reachable pump must trigger "pump unrechable" alert after
5 to 10 min.
- [ ] If the pump's profile doesn't match AAPS', the pump must be updated when AAPS starts
- [ ] Doing a profile change (to shift time or increase/decrease insulin), the pump's basal profile must be updated
- [ ] If a profile change has a duration, the pump's basal profile must be set to the original value again at the end
(this can vary a few minutes between what the overview screen shows and when the pump is updated, as the check
whether the pump is up-to-date or not is performed periodically and not at the exact minute a profile change ends)
- [ ] Read history using Smartpix and compare with AAPS' DB (treatment tab)
Esp. those times we communication was interrupted, boluses were cancelled, ...
- [ ] Boluses
- [ ] TBR
- [ ] Alerts
- [ ] TDDs
- [ ] Disconnected pump (pump unreachable)
- [ ] ...
- [ ] Refilling cartridge
- [ ] If TBR was cancelled by refilling, AAPS must detect this and update the TBR treatment
accordingly
- [ ] Stress testing
- PersistentNotification plugin disabled
- Lots of comms running, like Wifi, GSM, BT audio
- AAPS running in background
- Foreground app stresses the phone's memory, CPU (like a game) (potentially) pushing AAPS out of memory
- [ ] TBR must still be set/cancelled while running in the background
- [ ] With the pump powered off or out of reach, the 'pump unreachable alert' must still
trigger
- [ ] Combo tab
- [ ] Check displayed data (state, battery, reservoir, last bolus, temp basal) is the same
as on the pump
- [ ] Unsafe usage
- [ ] An extended or multiwave bolus given within the last six hour must raise an alert and
closed loop functionality (Loop plugin shows "disabled by constrains" when "execute" is pressed)
- [ ] Closed loop functionality must resume 6 h after the last ext/multiwave bolus
- [ ] An active ext/multiwave bolus (a history record is created only after the bolus completed)
must also raise an alert and disable loop
- [ ] If a basal rate other than profile 1 is activated, this must also raise an alert and disable
the loop for 6 h
- [ ] Reading/setting basal profile
- [ ] AAPS reads basal rate properly
- [ ] AAPS doesn't touch pump basal if "Sync to profile" pref not set
- [ ] AAPS updates basal rate if "Sync to profile" is enabled
- [ ] Test profile with 115% (or something like that) change to ask the
pump for basal rates like 0.812, which should then be set propely
- [ ] Taking over alerts
- [ ] If an error alert is active on the pump, pressing refresh shall display the error
in the Combo tab but NOT confirm it. Easiest error to trigger: rewind piston
and attempt to start the pump, this will trigger E11: Not primed.
- [ ] Pressing refresh while a low cartridge or low battery alarm is active
must confirm the alarm, indicate the new status in the Combo tab and
show a notification on the overview screen
- [ ] Misc
- [ ] Pump state is correctly uploaded to Nightscout