**This software is part of a DIY solution and is not a product, but
requires YOU to read, learn and understand the system and how to use it.
It is not something that does all your diabetes management for you, but
allows you to improve your diabetes and quality of life significantly
if you're willing to put in the time required.
You alone are responsible for what you do with it.**
Hardware requirements:
- A Roche Accu-Chek Combo (any firmware, they all work)
- A Smartpix or Realtyme device together with the 360 Configuration
Software to configure the pump.
Roche sends these out free of charge to their customers upon request.
- A compatible phone: An Android phone with a phone running LineageOS 14.1
(formerly CyanogenMod), or possibly stock Android >= 8.1 in the future.
Roche sends these out Smartpix devices and the configuration software
free of charge to their customers upon request.
- A compatible phone: An Android phone with a phone running LineageOS 14.1 (formerly CyanogenMod)
- To build AndroidAPS with Combo support you need the lastet Android Studio 3 version
Limitations:
- Extended bolus and multiwave bolus are not supported.
- Only one basal profile is supported.
- Setting a basal profile other than 1 on the pump, or delivering extended boluses or multiwave
boluses from the pump will disable the loop functionality for 6h as the the loop can't run
safely under those conditions.
- If multiple boluses are given within a single minute, only one will
be recognized. This is due to the Combo saving history records with
minute-precision only. However, this case is only possible for very
small boluses and is unlikely to occur in non-testing scenarios
(e.g. bolusing from the pump and then immediately bolusing from AAPS
or giving smaller boluses in the pump in quick succession).
- If a TBR is set on the pump, AAPS will cancel it. This is because it's not possible to determine
the start point of a TBR until it is finished or was cancelled at which point a record in the
pump's history is created. Before that, there is none and it's simply not possible to determine
the TBRs influence on IOB. Set TBR using AAPS instead.
- It's currently not possible to set the time and date on the pump (only reading the time - but
not the date - is supported raises a warning in AAPS to update the pump clock manually).
Thus, the pump's clock must be updated manually when the clocks are turned forward/backward
for daylight saving time.
boluses from the pump interferes with TBRs and forces the loop into low-suspend only mode for 6 hours
as the the loop can't run safely under those conditions.
- It's currently not possible to set the time and date on the pump, however, this has no effect
on AAPS and will not cause a problem if the pump's clock is not updated at the exact time
daylight savings time starts or ends.
- There's a bug in the pump's firmware that's triggered when "too much" communication happens
with the pump. Specifically, this issue occurs when going from just issuing commands to the pump
to reading the pumps data and history. For that reason, a minimal amount of data is read from
the pump (no history).
The bug might still rarely occur and causes the pump to not accept any connection
unless a button is physically pressed on the pump.
Therefore, the pump's reservoir level is not read and the pump status information uploaded to Nightscout
shows fake numbers of 250 (above low threshold - which can be configured via the configuration
tool), 50 (below low threshold) and 0 if the reservoir is empty.
Furthermore, no history (from the My Data menu) is read unless absolutely required.
Reading all pump data can be forced through the Combo tab (long press "TDDS"), to "import"
events that happend solely on the pump, but that code has been tested less and may
trigger the bug, so it's strongly recommended to stick to the usage scenario of controlling the
pump solely through AAPS.
Checking history, reservoir level etc on the pump causes no issues but should be avoided
when the Bluetooth icon is displayed on the display, indicating that AAPS is communicating with the pump.
Setup v2:
Setup:
- Configure pump using 360 config software.
- Set/leave the menu configuration as "Standard", this will show only the supported
menus/actions on the pump and hide those which are unsupported (extended/multiwave bolus,
@ -49,23 +57,18 @@ Setup v2:
can be controlled via ruffy, installing the above version is sufficient.
If AAPS is already installed, switch to the MDI plugin to avoid the Combo
plugin from interfering with ruffy during the pairing process.
- Get AndroidAPS from https://gitlab.com/jotomo/KEF (Branch `combo-scripter-v2`)
- Get AndroidAPS from https://github.com/jotomo/AndroidAPS (Branch `combo-scripter-v2`)
- Before enabling the Combo plugin in AAPS make sure you're profile is set up
correctly and your basal profile is up to date as AAPS will sync the basal profile
to the pump.
- There might be minor glitches around enabling/disabling the Combo plugi, requiring
to restart AAPS by force-closing it.
Usage:
- This is not a product, esp. in the beginning the user needs to monitor and understand the system,
its limitations and how it can fail. It is strongly advised NOT to use this system when the person
using is not able to fully understand the system.
- The integration of the Combo with AndroidAPS is designed with the assumption that all inputs are
made via AndroidAPS. While there are checks that will detected boluses entered directly on the
pump, which will be added to the history and be included in IOB calulations, there are delays
until AAPS becomes aware of those bolusus (up to 15m). It is therefore strongly adviced
to only bolus via AndroidAPS (it's also only possible to enter carbs via AndroidAPS, required for
advanced loop functionality).
made via AndroidAPS. Boluses entered on the pump will NOT be detected by AAPS and may therefore
result in too much insulin being delivered.
- It's recommended to enable key lock on the pump to prevent bolusing from the pump, esp. when the
pump was used before and quick bolusing was a habit.
Also, with keylock enabled, accidentally pressing a key will NOT interrupt a running command

View file

@ -1,5 +1,5 @@
- [ ] Bugs
- [ ] Pump stops to response to connect attemps every 2-20h or so, only reacting
- [-] Pump stops to response to connect attemps every 2-20h or so, only reacting
again when physically pressing a button on the pump ...
Occassionally the pump starts to accept connections again, sometimes
after 40m, sometimes no luck even after 80m
@ -19,7 +19,7 @@
Did this already happen before the whole 'basal profile' thing?
Takes ages, could that trigger some weird bug that has never
surfaced in any other scenario before?
- [ ] Temp basal history shows TBRs twice, once with a 1 min
- [-] Temp basal history shows TBRs twice, once with a 1 min
duration and then with the reminder; caused by clock not
being within ~30s of the phone.
Cancelling TBR or overriding a running one with a new one
@ -31,6 +31,9 @@
the Combo.
Explicit end records when adding pump history records;
and/or: logic to detect time differences and align data
Current solution works, albeit a bit ugly, but is used in rare
circumstances where clock might be off anyway, so this would
still occur.
- [x] Deleting a bolus from the history re-adds it from the pump's
history. Deleting it again, flags it as invalid at which point
it will not be added to IOB and not be re-added.

View file

@ -20,36 +20,30 @@
- [ ] Pressing a button on the pump after bolus delivery has started must be handled
gracefully (bolus finished and AAPS must correctly add it the the DB)
- [ ] Same as above, but moving pump out of range
- [ ] Giving a bolus on the pump and immediately afterwards entering a bolus on AAPS
must reject the bolus with a warning that IOB is not up to date
- [ ] A bolus given on the pump must appear in AAPS after the next communication or when
manually refreshing via the Combo tab
- [ ] Test the highest bolus you'd ever give yourself (AAPS has a configurable limit and the pump
has a limit which can be configured with the Config SW)
- [ ] Test a bolus exceeding the pumps max bolus; should show a generic error message
- [ ] BT disconnect issues
- [ ] Moving pump out of reach when setting TBR causes "TBR cancelled" alarm on pump.
When putting pump close to phone, AAPS must confirm the alert and successfully
retry setting TBR
retry setting TBR (reconnects are a best-effort kind of a thing, so this might not always work)
- [ ] Same as above while bolusing must report an error and NOT retry the command
- [ ] AAPS start
- [ ] Starting AAPS without a reachable pump must show something sensible in the Combo tab
(that is not hang indefinitely with "initializing" activity)
- [ ] Starting AAPS without a reachable pump must trigger "pump unrechable" alert after
5 to 10 min.
- [ ] If 'sync profile to pump' is set the basal rate profile must be synced to pump on start
- [ ] If 'sync profile to pump' is set NOT set, the pump's profile must not be touched
- [ ] If the pump's profile doesn't match AAPS', the pump must be updated when AAPS starts
- [ ] Doing a profile change (to shift time or increase/decrease insulin), the pump's basal profile must be updated
- [ ] If a profile change has a duration, the pump's basal profile must be set to the original value again at the end
(this can vary a few minutes between what the overview screen shows and when the pump is updated, as the check
whether the pump is up-to-date or not is performed periodically and not at the exact minute a profile change ends)
- [ ] Read history using Smartpix and compare with AAPS' DB (treatment tab)
Esp. those times we communication was interrupted, boluses were cancelled, ...
- [ ] Boluses
- [ ] TBR
- [ ] Alerts
- [ ] TDDs
- [ ] Date/Time
- [ ] If time is off by more than 2m, a warning must be shown on the main screen whenever
a connection to the pump is established
- [ ] Since history records store the year, it must be inferred. Set phone and pump to something
like 31.21.17, 23:00, cause an alert, deliver a bolus, set a TBR, let the new year arrive
trigger the above events again and verify AAPS has correctly read them.
- [ ] Disconnected pump (pump unreachable)
- [ ] ...
- [ ] Refilling cartridge
@ -66,8 +60,6 @@
- [ ] Combo tab
- [ ] Check displayed data (state, battery, reservoir, last bolus, temp basal) is the same
as on the pump
- [ ] Last connection must indicate issues (yellow if recent connections/commands failed),
red if communications failed for 30 or more minutes
- [ ] Unsafe usage
- [ ] An extended or multiwave bolus given within the last six hour must raise an alert and
closed loop functionality (Loop plugin shows "disabled by constrains" when "execute" is pressed)