- Bolusing
- Cancelling bolus at various stages and checking error message (if any) and check no bolus or a partial bolus was delivered
- Cancelling a bolus 0.2 U before its finished (cancelling requires pressing up/down 3s, so the cancellation will be too late, but the code should handle it and report the full bolus being delivered
- Low cartridge alarm during bolus
- must be confirmed by AAPS
- bolus must have been fully delivered by pump
- bolus must have been added to DB
- the confirmed pump warning must be raised as a notification in AAPS
- notification must also appear on smartwatch
- An error during bolus must yield an error in AAPS
- An error during bolus must yield a notification on a smartwatch
- Bolusing e.g. 4 U if reservoir has only 2 U left must be rejected
- Pressing a button on the pump before bolus delivery started must be handled gracefully
- Same as above, but moving pump out of range
- Pressing a button on the pump after bolus delivery has started must be handled
gracefully (bolus finished and AAPS must correctly add it the the DB)
- Same as above, but moving pump out of range
- Giving a bolus on the pump and immediately afterwards entering a bolus on AAPS must reject the bolus with a warning that IOB is not up to date
- A bolus given on the pump must appear in AAPS after the next communication or when manually refreshing via the Combo tab
- Test the highest bolus you'd ever give yourself (AAPS has a configurable limit and the pump has a limit which can be configured with the Config SW)
- BT disconnect issues
- Moving pump out of reach when setting TBR causes "TBR cancelled" alarm on pump. When putting pump close to phone, AAPS must confirm the alert and successfully retry setting TBR
- Same as above while bolusing must report an error and NOT retry the command
- AAPS start
- Starting AAPS without a reachable pump must show something sensible in the Combo tab (that is not hang indefinitely with "initializing" activity)
- Starting AAPS without a reachable pump must trigger "pump unrechable" alert after 5 to 10 min.
- If 'sync profile to pump' is set the basal rate profile must be synced to pump on start
- If 'sync profile to pump' is set NOT set, the pump's profile must not be touched
- Read history using Smartpix and compare with AAPS' DB (treatment tab)
Esp. those times we communication was interrupted, boluses were cancelled, ...
- Boluses
- TBR
- Alerts
- TDDs
- Date/Time
- If time is off by more than 2m, a warning must be shown on the main screen whenever a connection to the pump is established
- Since history records store the year, it must be inferred. Set phone and pump to something like 31.21.17, 23:00, cause an alert, deliver a bolus, set a TBR, let the new year arrive trigger the above events again and verify AAPS has correctly read them.
- Disconnected pump (pump unreachable) - [ ] ...
- Refilling cartridge
- If TBR was cancelled by refilling, AAPS must detect this and update the TBR treatment accordingly
- Stress testing
- PersistentNotification plugin disabled
- Lots of comms running, like Wifi, GSM, BT audio
- AAPS running in background
- Foreground app stresses the phone's memory, CPU (like a game) (potentially) pushing AAPS out of memory
- TBR must still be set/cancelled while running in the background
- With the pump powered off or out of reach, the 'pump unreachable alert' must still trigger
- Combo tab
- Check displayed data (state, battery, reservoir, last bolus, temp basal) is the same as on the pump
- Last connection must indicate issues (yellow if recent connections/commands failed), red if communications failed for 30 or more minutes
- Unsafe usage
- An extended or multiwave bolus given within the last six hour must raise an alert and closed loop functionality (Loop plugin shows "disabled by constrains" when "execute" is pressed)
- Closed loop functionality must resume 6 h after the last ext/multiwave bolus
- An active ext/multiwave bolus (a history record is created only after the bolus completed) must also raise an alert and disable loop
- If a basal rate other than profile 1 is activated, this must also raise an alert and disable the loop for 6 h
- Reading/setting basal profile
- AAPS reads basal rate properly
- AAPS doesn't touch pump basal if "Sync to profile" pref not set
- AAPS updates basal rate if "Sync to profile" is enabled
- Test profile with 115% (or something like that) change to ask the pump for basal rates like 0.812, which should then be set propely
- Taking over alerts
- If an error alert is active on the pump, pressing refresh shall display the error in the Combo tab but NOT confirm it. Easiest error to trigger: rewind piston and attempt to start the pump, this will trigger E11: Not primed.
- Pressing refresh while a low cartridge or low battery alarm is active must confirm the alarm, indicate the new status in the Combo tab and show a notification on the overview screen