**This software is part of a DIY solution and is not a product, but
|
|
requires YOU to read, learn and understand the system and how to use it.
|
|
It is not something that does all your diabetes management for you, but
|
|
allows you to improve your diabetes and quality of life significantly
|
|
if you're willing to put in the time required. Don't rush into it,
|
|
but allow yourself time to learn. You alone are responsible for what
|
|
you do with it.**
|
|
|
|
Hardware requirements
|
|
---------------------
|
|
- A Roche Accu-Chek Combo (any firmware, they all work)
|
|
- A Smartpix or Realtyme device together with the 360 Configuration
|
|
Software to configure the pump.
|
|
Roche sends out Smartpix devices and the configuration software
|
|
free of charge to their customers upon request.
|
|
- A compatible phone: An Android phone with a phone running LineageOS 14.1 (formerly CyanogenMod) or Android 8.1 (Oreo).
|
|
Be aware that while Android 8.1 allows communicating with the Combo, there are still issues with AAPS on 8.1.
|
|
For advanced users, it is possible to perform the pairing on a rooted phone and transfer it to another rooted
|
|
phone to use with ruffy/AAPS, which must also be rooted. This allows using phones with Android < 8.1 but
|
|
has not been widely tested: https://github.com/gregorybel/combo-pairing/blob/master/README.md
|
|
- To build AndroidAPS with Combo support you need the latest Android Studio 3 version
|
|
|
|
Limitations
|
|
-----------
|
|
- Extended bolus and multiwave bolus are not supported.
|
|
- Only one basal profile is supported.
|
|
- Setting a basal profile other than 1 on the pump, or delivering extended boluses or multiwave
|
|
boluses from the pump interferes with TBRs and forces the loop into low-suspend only mode for 6 hours
|
|
as the the loop can't run safely under those conditions.
|
|
- It's currently not possible to set the time and date on the pump, so daylight saving times
|
|
changes have to be performed manually (disable automatic phone clock update in the evening and
|
|
update change back in the morning while also updating the pump's clock to avoid an alarm during the night).
|
|
- Currently only basal rates in the range of 0.05 to 10 U/h are supported (this also applies when modifying
|
|
a profile, e.g. when increasing to 200%, the highest basal rate must not exceed 5 U/h since it will be
|
|
doubled. Similarly, when reducing to 50%, the lowest basal rate must be at least 0.10 U/h).
|
|
- If the loop requests a running TBR to be cancelled the Combo will set a TBR of 90% or 110%
|
|
for 15 minutes instead. This is because cancelling a TBR causes an alert on the pump which
|
|
causes a lot of vibrations.
|
|
- Occasionally (every couple of days or so) AAPS might fail to automatically cancel
|
|
a TBR CANCELLED alert the user then needs to deal with (press the refresh button in AAPS
|
|
to transfer the warning to AAPS or confirm the alert on the pump).
|
|
|
|
Setup
|
|
-----
|
|
- Configure pump using 360 config software.
|
|
- Required:
|
|
- Set/leave the menu configuration as "Standard", this will show only the supported
|
|
menus/actions on the pump and hide those which are unsupported (extended/multiwave bolus,
|
|
multiple basal rates), which cause the loop functionality to be restricted when used because
|
|
it's not possible to run the loop in a safe manner when used.
|
|
- Verify the _Quick Info Text_ is set to "QUICK INFO" (without the quotes, found under _Insulin Pump Options_).
|
|
- Set maximum TBR to 500%
|
|
- Disable end of TBR alert
|
|
- Set TBR duration step-size to 15 min
|
|
- Recommended:
|
|
- Set low cartridge alarm to your liking
|
|
- Configure a max bolus suited for your therapy to protect against bugs in the software
|
|
- Similarly, configure maximum TBR duration as a safeguard. Allow at least 3 hours, since
|
|
the option to disconnect the pump for 3 hours sets a 0% for 3 hours.
|
|
- Enable key lock on the pump to prevent bolusing from the pump, esp. when the
|
|
pump was used before and quick bolusing was a habit.
|
|
- Set display timeout and menu timeout to the mininum of 5.5 and 5 respectively. This allows the AAPS to
|
|
recover more quickly from error situations and reduces the amount of vibrations that can occur during
|
|
such errors
|
|
- Install AndroidAPS as described in the wiki http://wiki.AndroidAPS.org
|
|
- Make sure to read the wiki to understand how to setup AndroidAPS.
|
|
- Select the MDI plugin in AndroidAPS, not the Combo plugin at this point to avoid the Combo
|
|
plugin from interfering with ruffy during the pairing process.
|
|
- Follow the link http://ruffy.AndroidAPS.org and clone ruffy via git. Use the same branch as you use for
|
|
AndroidAPS, which is usally `master`.
|
|
if you you want to help test the in-development version.
|
|
- Pair the pump using ruffy. If it doesn't work after multiple attempts, switch to the `pairing` branch,
|
|
pair the pump and then switch back the original branch.
|
|
If the pump is already paired and can be controlled via ruffy, installing the `master` branch is sufficient.
|
|
Note that the pairing processing is somewhat fragile (but only has to be done once)
|
|
and may need a few attempts; quickly acknowledge prompts and when starting over, remove the pump device
|
|
from the bluetooth settings beforehand. Another option to try is to go to the bluetooth menu after
|
|
initiating the pairing process (this keeps the phone's bluetooth discoverable as long as the menu is displayed)
|
|
and switch back to ruffy after confirming the pairing on the pump, when the pump displays the authorization code.
|
|
When AAPS is using ruffy, the ruffy app can't be used. The easiest way is to just
|
|
reboot the phone after the pairing process and let AAPS start ruffy in the background.
|
|
- Before enabling the Combo plugin in AAPS make sure your profile is set up
|
|
correctly and your basal profile is up to date as AAPS will sync the basal profile
|
|
to the pump. Then activate the Combo plugin.
|
|
|
|
Usage
|
|
-----
|
|
- Keep in mind that this is not a product, esp. in the beginning the user needs to monitor and understand the system,
|
|
its limitations and how it can fail. It is strongly advised NOT to use this system when the person
|
|
using it is not able to fully understand the system.
|
|
- Read the OpenAPS documentation https://openaps.org to understand the loop algorithm AndroidAPS
|
|
is based upon.
|
|
- Read in the wiki to learn about and understand AndroidAPS http://wiki.AndroidAPS.org
|
|
- This integration uses the same functionality which the meter provides that comes with the Combo.
|
|
The meter allows to mirror the pump screen and forwards button presses to the pump. The connection
|
|
to the pump and this forwarding is what the ruffy app does. A `scripter` components reads the screen
|
|
and automates inputing boluses, TBRs etc and making sure inputs are processed correctly.
|
|
AAPS then interacts with the scripter to apply loop commands and to administer boluses.
|
|
This mode has some restrictions: it's comparatively slow (but well fast enough for what it is used for),
|
|
and setting a TBR or giving a bolus causes the pump to vibrate.
|
|
- The integration of the Combo with AndroidAPS is designed with the assumption that all inputs are
|
|
made via AndroidAPS. Boluses entered on the pump directly will be detected by AAPS, but it can take
|
|
up to 25 min before AndroidAPS becomes aware of such a bolus. Reading boluses delivered directly on
|
|
the pump is a safety feature and not meant to be regularly used (the loop requires knowledge of carbs
|
|
consumed, which can't be entered on the pump, which is another reason why all inputs should be done
|
|
in AndroidAPS).
|
|
- The pump's first basal rate profile is read on application start and is updated by AAPS.
|
|
The basal rate should not be manually changed on the pump, but will be detected and corrected as a safety
|
|
measure (don't rely on safety measures by default, this is meant to detect an unintended change on the pump).
|
|
- It's recommended to enable key lock on the pump to prevent bolusing from the pump, esp. when the
|
|
pump was used before and quick bolusing was a habit.
|
|
Also, with keylock enabled, accidentally pressing a key will NOT interrupt active communication
|
|
between AAPS and pmup.
|
|
- When a BOLUS/TBR CANCELLED alert starts on the pump during bolusing or setting a TBR, this is
|
|
caused by a disconnect between pump and phone. AAPS will try to reconnect and confirm the alert
|
|
and then retry the last action (boluses are NOT retried for safety reasons). Therefore,
|
|
such an alarm shall be ignored (cancelling it is not a big issue, but will lead to the currently
|
|
active action to have to wait till the pump's display turns off before it can reconnect to the
|
|
pump). If the pump's alarm continues, the last action might have failed, in which case the user
|
|
needs to confirm the alarm.
|
|
- When a low cartridge or low battery alarm is raised during a bolus, they are confirmed and shown
|
|
as a notification in AAPS. If they occur while no connection is open to the pump, going to the
|
|
Combo tab and hitting the Refresh button will take over those alerts by confirming them and
|
|
showing a notification in AAPS.
|
|
- When AAPS fails to confirm a TBR CANCELLED alert, or one is raised for a different reason,
|
|
hitting Refresh in the Combo tab establishes a connection, confirms the alert and shows
|
|
a notification for it in AAPS. This can safely be done, since those alerts are benign - an
|
|
appropriate TBR will be set again during the next loop iteration.
|
|
- For all other alerts raised by the pump: connecting to the pump will show the alert message in
|
|
the Combo tab, e.g. "State: E4: Occlusion" as well as showing a notification on the main screen.
|
|
An error will raise an urgent notification. AAPS never confirms serious errors on the pump,
|
|
but let's the pump vibrate and ring to make sure the user is informed of a critical situation
|
|
that needs action.
|
|
- After pairing, ruffy should not be used directly (AAPS will start in the background as needed),
|
|
since using ruffy at AAPS at the same time is not supported.
|
|
- If AAPS crashes (or is stopped from the debugger) while AAPS and the pump were communicating (using
|
|
ruffy), it might be necessary to force close ruffy. Restarting AAPS will start ruffy again.
|
|
Restarting the phone is also an easy way to resolve this or you don't know how to force kill
|
|
an app.
|
|
- Don't press any buttons on the pump while AAPS communicates with the pump (Bluetooth logo is
|
|
shown on the pump).
|
|
|
|
Reporting bugs
|
|
--------------
|
|
- See https://github.com/MilosKozak/AndroidAPS/blob/master/ISSUE_TEMPLATE.md
|
|
- Note the precise time the problem occurred and describe the circumstances and steps that caused
|
|
the problem
|
|
- Note the Build version (found in the About dialog in the app, when pressing the three dots in the
|
|
upper-right corner).
|
|
- Obtain the app's log files, which can be found on the phone in
|
|
_/storage/emulated/0/Android/data/info.nightscout.androidaps/_
|
|
- Open an issue at https://github.com/MilosKozak/AndroidAPS/issues/new
|
|
|