AndroidAPS/Testing-Combo.md
2017-12-11 20:44:29 +01:00

97 lines
6.4 KiB
Markdown

- [ ] Pairing
- [ ] Pairing works with `combo-scripter-v2` branch
- [ ] Bolusing
- [ ] Cancelling bolus at various stages and checking error message (if any) and check
no bolus or a partial bolus was delivered
- [ ] Enter a bolus of 2 U and press cancel when delivery is at 1.7 (cancelling requires AAAPS
to press the up button for 3 seconds, so the cancellation attempt will not succeed because delivery
ends before those 3 seconds are elapsed). The code should handle this without giving an
error and add the full bolus to treatments.
- [ ] Low cartridge alarm during bolus
- [ ] alarm must be confirmed by AAPS
- [ ] bolus must have been fully delivered by pump
- [ ] bolus must have been added to DB
- [ ] the confirmed pump warning must be raised as a notification in AAPS
- [ ] notification must also appear on smartwatch
- [ ] An error during bolus must yield an error in AAPS
- [ ] An error during bolus must yield a notification on a smartwatch
- [ ] Test bolusing a bolus bigger than what's left in the reservoir. A message to check what
was actually delivered must appear (this is a corner-case where we practically can't
check what was actually delivered).
- [ ] Pressing a button on the pump before bolus delivery started must be handled gracefully
- [ ] Same as above, but moving pump out of range
- [ ] Pressing a button on the pump after bolus delivery has started must raise a warning
to check what was actually delivered and manually update treatments in AAPS.
- [ ] Same as above, but moving pump out of range
- [ ] Test the highest bolus you'd ever give yourself (AAPS has a configurable limit and the pump
has a limit which can be configured with the Config SW), no timeout or other issues must show
- [ ] Test a bolus exceeding the pumps max bolus; should show a generic error message
- [ ] BT disconnect issues
- [ ] Moving pump out of reach when setting TBR causes "TBR cancelled" alarm on pump.
When putting pump close to phone, AAPS must confirm the alert and successfully
retry setting TBR (reconnects are a best-effort kind of a thing, so this might not always work)
- [ ] When a disconnect occurs, the pump's screen shows the error and the pump only accepts a connection
again when a timeout has occurred. A recovery should be quicker if that timeout is decreased.
It might be interesting to experiment with the Config software to set lower menu or display timeouts
(or whatever they're called ...) to improve recovery speed.
- [ ] Same as above while bolusing must report an error and NOT retry the command
- [ ] AAPS start
- [ ] Starting AAPS without a reachable pump must show something sensible in the Combo tab
(not hanging indefinitely with "initializing" activity)
- [ ] Starting AAPS without a reachable pump must trigger "pump unrechable" alert after the configured threshold
- [ ] If the pump's profile doesn't match AAPS', the pump must be updated when AAPS starts
- [ ] Read history using Smartpix and compare with AAPS' DB (treatment tab)
Esp. those times we communication was interrupted, boluses were cancelled, ...
- [ ] Boluses
- [ ] TBR
- [ ] Alerts
- [ ] Disconnected pump (pump unreachable)
- [ ] With local alerts enabled for 'pump unreachable', an alert must be triggered within 5 minutes
after the configured threshold. (Don't set the threshold too low, e.g. 10 minutes, since
there might be no need to set a TBR within such a short time).
- [ ] Refilling cartridge
- [ ] If TBR was cancelled by refilling, AAPS must detect this and create a TBR record in AAPS
based on what the pump displays (not the full TBR duration, but what is displayed as remaining
on the main screen.
- [ ] Stress testing
- PersistentNotification plugin disabled
- Lots of comms running, like Wifi, GSM, BT audio
- AAPS running in background
- Foreground app stresses the phone's memory, CPU (like a game) (potentially) pushing AAPS out of memory
- [ ] TBR must still be set/cancelled while running in the background
- [ ] With the pump powered off or out of reach, the 'pump unreachable alert' must still
trigger
- [ ] Combo tab
- [ ] Check displayed data (state, battery, reservoir, temp basal) is the same
as on the pump
- [ ] Unsafe usage
- [ ] An extended or multiwave bolus given within the last six hour must raise an alert and
restrict the loop functionality to low-suspend only (setting maxIOB to zero)
- [ ] Closed loop functionality must resume 6 h after the last ext/multiwave bolus
- [ ] An active ext/multiwave bolus must also raise an alert and restrict the loop
- [ ] If a basal rate other than profile 1 is activated, this must also raise an alert and disable
the restrict the loop
- [ ] Reading/setting basal profile
- [ ] AAPS reads basal rate properly
- [ ] Test profile with 115% (or something like that) change to ask the
pump for basal rates like 0.812, which should then be set propely
- [ ] Updating the profile extensively (200%, shifting time) takes up to 6 minutes, but
should complete without timeout.
- [ ] Doing a profile change (to shift time or increase/decrease insulin), the pump's basal profile must be updated
- [ ] If a profile change has a duration, the pump's basal profile must be set to the original value again at the end
(this can vary a few minutes between what the overview screen shows and when the pump is updated, as the check
whether the pump is up-to-date or not is performed periodically and not at the exact minute a profile change ends)
- [ ] Taking over alerts
- [ ] If an error alert is active on the pump, pressing refresh shall display the error
in the Combo tab but NOT confirm it. Easiest error to trigger: rewind piston
and attempt to start the pump, this will trigger E11: Not primed.
- [ ] Pressing refresh while a low cartridge or low battery alarm is active
must confirm the alarm, indicate the new status in the Combo tab and
show a notification on the overview screen
- [ ] A TBR CANCELLED is now also taken over when refreshing (since it's a benign error the loop will correct
during the next iteration).
- [ ] Misc
- [ ] Pump state is correctly uploaded to Nightscout (note that reservoir level are fake numbers representing
norma/low/empty).