This software is part of a DIY solution and is not a product, but requires YOU to read, learn and understand the system and how to use it. It is not something that does all your diabetes management for you, but allows you to improve your diabetes and quality of life significantly if you're willing to put in the time required. You alone are responsible for what you do with it.
Hardware requirements:
- A Roche Accu-Chek Combo (any firmware, they all work)
- A Smartpix or Realtyme device together with the 360 Configuration Software to configure the pump. Roche sends out Smartpix devices and the configuration software free of charge to their customers upon request.
- A compatible phone: An Android phone with a phone running LineageOS 14.1 (formerly CyanogenMod)
- To build AndroidAPS with Combo support you need the latest Android Studio 3 version
Limitations:
- Extended bolus and multiwave bolus are not supported.
- Only one basal profile is supported.
- Setting a basal profile other than 1 on the pump, or delivering extended boluses or multiwave boluses from the pump interferes with TBRs and forces the loop into low-suspend only mode for 6 hours as the the loop can't run safely under those conditions.
- It's currently not possible to set the time and date on the pump, so daylight saving times changes have to be performed manually.
- There's a bug in the pump's firmware that's triggered when "too much" communication happens with the pump. Specifically, this issue occurs when going from just issuing commands to the pump to reading the pump's data and history. For that reason, a minimal amount of data is read from the pump. The bug might still rarely occur and causes the pump to not accept any connection unless a button is physically pressed on the pump (make sure the 'pump unreachable' alert is enabled so you'll get an alarm if the bug occurs. Therefore, the pump's reservoir level is not read and the pump status information uploaded to Nightscout shows fake numbers of 150 (above low threshold - which can be configured via the configuration tool), 8 (below low threshold, triggers NS alarm since < 10) and 0 if the reservoir is empty. Furthermore, no history (from the My Data menu) is read unless absolutely required. Reading all pump data can be forced through the Combo tab, to import events that happened solely on the pump, but that code has been tested less, might cause duplicates and may trigger the bug, so it's strongly recommended to stick to the usage scenario of controlling the pump solely through AAPS. Checking history, reservoir level etc on the pump causes no issues but should be avoided when the Bluetooth icon is displayed on the display, indicating that AAPS is communicating with the pump.
- Currently only basal rates in the range of 0.05 to 10 U/h are supported (this also applies when modifying a profile, e.g. when increasing to 200%, the highest basal rate must not exceed 5 U/h since it will be doubled. Similarly, when reducing to 50%, the lowest basal rate must be at least 0.10 U/h).
Setup:
- Configure pump using 360 config software.
- Set/leave the menu configuration as "Standard", this will show only the supported menus/actions on the pump and hide those which are unsupported (extended/multiwave bolus, multiple basal rates), which cause the loop functionality to be restricted when used because it's not possible to run the loop in a safe manner when used.
- Set maximum TBR to 500%
- Disable end of TBR alert
- Set TBR duration step-size to 15 min
- Set low cartridge alarm to your liking
- Enable keylock (can also be set on the pump directly, see usage section on reasoning)
- Get Android Studio 3 https://developer.android.com/studio/index.html
- Clone ruffy from https://github.com/jotomo/ruffy (branch
combo-scripter-v2
) - Pair the pump, if it doesn't work after multiple attempts, switch to the
pairing
branch, pair, then switch back the original branch. If the pump is already paired and can be controlled via ruffy, installing the above version is sufficient. If AAPS is already installed, switch to the MDI plugin to avoid the Combo plugin from interfering with ruffy during the pairing process. Note that the pairing processing is somewhat fragile and may need a few attempts; quickly acknowledge prompts and when starting over, remove the pump device from the bluetooth settings beforehand. - Clone AndroidAPS from https://github.com/jotomo/AndroidAPS (Branch
combo-scripter-v2
) - Before enabling the Combo plugin in AAPS make sure your profile is set up correctly and your basal profile is up to date as AAPS will sync the basal profile to the pump.
Usage:
- Keep in mind that this is not a product, esp. in the beginning the user needs to monitor and understand the system, its limitations and how it can fail. It is strongly advised NOT to use this system when the person using is not able to fully understand the system.
- This integration uses the same functionality which the meter provides that comes with the Combo.
The meter allows to mirror the pump screen and forwards button presses to the pump. The connection
to the pump and this forwarding is what the ruffy app does. A
scripter
components reads the screen and automates inputing boluses, TBRs etc and making sure inputs are processed correctly (that's what the scripter-part in the branch name stands for). AAPS then interacts with the scripter to apply loop commands and to administer boluses. This mode has some restrictions: it's comparatively slow (but well fast enough for what it is used for), doesn't support reading history continuously and setting a TBR or giving a bolus causes the pump to vibrate. - The integration of the Combo with AndroidAPS is designed with the assumption that all inputs are made via AndroidAPS. Boluses entered on the pump will NOT be detected by AAPS and may therefore result in too much insulin being delivered.
- It's recommended to enable key lock on the pump to prevent bolusing from the pump, esp. when the pump was used before and quick bolusing was a habit. Also, with keylock enabled, accidentally pressing a key will NOT interrupt a running command
- When a BOLUS/TBR CANCELLED alert starts on the pump during bolusing or setting a TBR, this is caused by a disconnect between pump and phone. The app will try to reconnect and confirm the alert and then retry the last action (boluses are NOT retried for safety reasons). Therefore, such an alarm shall be ignored (cancelling it is not a big issue, but will lead to the currently active action to have to wait till the pump's display turns off before it can reconnect to the pump). If the pump's alarm continues, the last action might have failed, in which case the user needs to confirm the alarm.
- When a low cartridge or low battery alarm is raised during a bolus, they are confirmed and shown as a notification in AAPS. If they occur while no connection is open to the pump, going to the combo tab and hitting the Refresh button will take over those alerts by confirming them and showing a notification in AAPS.
- When AAPS fails to confirm a TBR CANCELLED alert, or one is raised for a different reason, hitting Refresh in the Combo tab establishes a connection, confirms the alert and shows a notification for it in AAPS. This can safely be done, since those alerts are benign - an appropriate TBR will be set again during the next loop iteration.
- For all other alerts raised by the pump: connecting to the pump will show the alert message in the Combo tab, e.g. "State: E4: Occlusion" as well as showing a notification on the main screen. An error will raise an urgent notification.
- After pairing, ruffy should not be used directly (AAPS will start in the background as needed), since using ruffy at AAPS at the same time is not supported.
- If AAPS crashes (or is stopped from the debugger) while AAPS and the pump were comunicating (using ruffy), it might be necessary to force close ruffy. Restarting AAPS will start ruffy again.
- Read the documentation on the wiki as well as the docs at https://openaps.org
- Don't press any buttons on the pump while AAPS communicates with the pump (Bluetooth logo is shown on the pump).
- If the loop requests a running TBR to be cancelled the Combo will set a TBR of 90% or 110% for 15 minutes instead. This is because cancelling a TBR causes an alert on the pump which causes a lot of vibrations.
- Due to the bug (which causes the pump to become unreachable when reading history regularly), a delivered bolus will NOT be added to treatments when the connection was lost during bolusing or when a pump error occurs (e.g. occlusion). This will raise a message (and play an annoying sound) and the user will have a create a bolus record via Careportal/NS manually. Currently this requires an available NS installation and being online (for a short time). Hopefully both of these issues can be resolved in future versions.
Known issues:
- Occasionally (every couple of days or so) AAPS might fail to automatically cancel a TBR CANCELLED alert the user then needs to deal with (press the refresh button in AAPS to transfer the warning to AAPS or confirm the alert on the pump).
- Similarly, the 'pump unreachable' bug may occur from time to time, which requires confirming the alert on the pump to get the pump to accept connections again.
- Overall the integration seems rather robust, but there are limits to the way the pump is controlled and how stable BT is, so there will be minor issues like the above from time to time, though they're small compared to what works well.
Reporting bugs:
- Note the precise time the problem occurred and describe the circumstances and steps that caused the problem
- Note the Build version (found in the About dialog in the app, when pressing the three dots in the upper-right corner).
- Attach the app's log files, which can be found on the phone in /storage/emulated/0/Android/data/info.nightscout.androidaps/
Components/Layers (developers):
- AndroidAPS
- ComboPlugin
- ruffy-spi (RuffyCommands interface to plug in lower layer)
- Scripting layer (ruffyscripter) / Command layer
- Driver layer (ruffy)