**This software is part of a DIY solution and is not a product, but
|
|
requires YOU to read, learn and understand the system and how to use it.
|
|
It is not something that does all your diabetes management for you, but
|
|
allows you to improve your diabetes and quality of life significantly
|
|
if you're willing to put in the time required.
|
|
You alone are responsible for what you do with it.**
|
|
|
|
Hardware requirements:
|
|
- A Roche Accu-Chek Combo (any firmware, they all work)
|
|
- A Smartpix or Realtyme device together with the 360 Configuration
|
|
Software to configure the pump.
|
|
Roche sends out Smartpix devices and the configuration software
|
|
free of charge to their customers upon request.
|
|
- A compatible phone: An Android phone with a phone running LineageOS 14.1 (formerly CyanogenMod)
|
|
- To build AndroidAPS with Combo support you need the latest Android Studio 3 version
|
|
|
|
Limitations:
|
|
- Extended bolus and multiwave bolus are not supported.
|
|
- Only one basal profile is supported.
|
|
- Setting a basal profile other than 1 on the pump, or delivering extended boluses or multiwave
|
|
boluses from the pump interferes with TBRs and forces the loop into low-suspend only mode for 6 hours
|
|
as the the loop can't run safely under those conditions.
|
|
- It's currently not possible to set the time and date on the pump, so daylight saving times
|
|
changes have to be performed manually.
|
|
- There's a bug in the pump's firmware that's triggered when "too much" communication happens
|
|
with the pump. Specifically, this issue occurs when going from just issuing commands to the pump
|
|
to reading the pump's data and history. For that reason, a minimal amount of data is read from
|
|
the pump.
|
|
The bug might still rarely occur and causes the pump to not accept any connection
|
|
unless a button is physically pressed on the pump (make sure the 'pump unreachable' alert is enabled
|
|
so you'll get an alarm if the bug occurs.
|
|
Therefore, the pump's reservoir level is not read and the pump status information uploaded to Nightscout
|
|
shows fake numbers of 150 (above low threshold - which can be configured via the configuration
|
|
tool), 8 (below low threshold, triggers NS alarm since < 10) and 0 if the reservoir is empty.
|
|
Furthermore, no history (from the My Data menu) is read unless absolutely required.
|
|
Reading all pump data can be forced through the Combo tab, to import
|
|
events that happened solely on the pump, but that code has been tested less, might cause duplicates and may
|
|
trigger the bug, so it's strongly recommended to stick to the usage scenario of controlling the
|
|
pump solely through AAPS.
|
|
Checking history, reservoir level etc on the pump causes no issues but should be avoided
|
|
when the Bluetooth icon is displayed on the display, indicating that AAPS is communicating with the pump.
|
|
- Currently only basal rates in the range of 0.05 to 10 U/h are supported (this also applies when modifying
|
|
a profile, e.g. when increasing to 200%, the highest basal rate must not exceed 5 U/h since it will be
|
|
doubled. Similarly, when reducing to 50%, the lowest basal rate must be at least 0.10 U/h).
|
|
|
|
Setup:
|
|
- Configure pump using 360 config software.
|
|
- Set/leave the menu configuration as "Standard", this will show only the supported
|
|
menus/actions on the pump and hide those which are unsupported (extended/multiwave bolus,
|
|
multiple basal rates), which cause the loop functionality to be restricted when used because
|
|
it's not possible to run the loop in a safe manner when used.
|
|
- Set maximum TBR to 500%
|
|
- Disable end of TBR alert
|
|
- Set TBR duration step-size to 15 min
|
|
- Set low cartridge alarm to your liking
|
|
- Enable keylock (can also be set on the pump directly, see usage section on reasoning)
|
|
- Get Android Studio 3 https://developer.android.com/studio/index.html
|
|
- Follow the link http://ruffy.AndroidAPS.org and clone via git (branch `combo-scripter-v2`)
|
|
- Pair the pump using ruffy, if it doesn't work after multiple attempts, switch to the `pairing` branch, pair,
|
|
then switch back the original branch. If the pump is already paired and
|
|
can be controlled via ruffy, installing the above version is sufficient.
|
|
If AAPS is already installed, switch to the MDI plugin to avoid the Combo
|
|
plugin from interfering with ruffy during the pairing process.
|
|
Note that the pairing processing is somewhat fragile (but only has to be done once)
|
|
and may need a few attempts;
|
|
quickly acknowledge prompts and when starting over, remove the pump device
|
|
from the bluetooth settings beforehand.
|
|
When AAPS is using ruffy, the ruffy app can't be used. The easiest way is to just
|
|
reboot the phone after the pairing process and let AAPS start ruffy in the background.
|
|
- Clone AndroidAPS from https://github.com/jotomo/AndroidAPS (branch `combo-scripter-v2`)
|
|
and build AAPS using the instructions in the wiki http://wiki.AndroidAPS.org
|
|
- Before enabling the Combo plugin in AAPS make sure your profile is set up
|
|
correctly and your basal profile is up to date as AAPS will sync the basal profile
|
|
to the pump.
|
|
|
|
Usage:
|
|
- Keep in mind that this is not a product, esp. in the beginning the user needs to monitor and understand the system,
|
|
its limitations and how it can fail. It is strongly advised NOT to use this system when the person
|
|
using is not able to fully understand the system.
|
|
- This integration uses the same functionality which the meter provides that comes with the Combo.
|
|
The meter allows to mirror the pump screen and forwards button presses to the pump. The connection
|
|
to the pump and this forwarding is what the ruffy app does. A `scripter` components reads the screen
|
|
and automates inputing boluses, TBRs etc and making sure inputs are processed correctly (that's what
|
|
the scripter-part in the branch name stands for).
|
|
AAPS then interacts with the scripter to apply loop commands and to administer boluses.
|
|
This mode has some restrictions: it's comparatively slow (but well fast enough for what it is used for),
|
|
doesn't support reading history continuously and setting a TBR or giving a bolus causes the pump to
|
|
vibrate.
|
|
- The integration of the Combo with AndroidAPS is designed with the assumption that all inputs are
|
|
made via AndroidAPS. Boluses entered on the pump will NOT be detected by AAPS and may therefore
|
|
result in too much insulin being delivered.
|
|
- The pump's first basal rate profile is read on app start and is updated by AAPS. Manually changing
|
|
the pump's basal rate profile will lead to wrong basals being delivered and is NOT supported.
|
|
- It's recommended to enable key lock on the pump to prevent bolusing from the pump, esp. when the
|
|
pump was used before and quick bolusing was a habit.
|
|
Also, with keylock enabled, accidentally pressing a key will NOT interrupt a running command
|
|
- When a BOLUS/TBR CANCELLED alert starts on the pump during bolusing or setting a TBR, this is
|
|
caused by a disconnect between pump and phone. The app will try to reconnect and confirm the alert
|
|
and then retry the last action (boluses are NOT retried for safety reasons). Therefore,
|
|
such an alarm shall be ignored (cancelling it is not a big issue, but will lead to the currently
|
|
active action to have to wait till the pump's display turns off before it can reconnect to the
|
|
pump). If the pump's alarm continues, the last action might have failed, in which case the user
|
|
needs to confirm the alarm.
|
|
- When a low cartridge or low battery alarm is raised during a bolus, they are confirmed and shown
|
|
as a notification in AAPS. If they occur while no connection is open to the pump, going to the
|
|
combo tab and hitting the Refresh button will take over those alerts by confirming them and
|
|
showing a notification in AAPS.
|
|
- When AAPS fails to confirm a TBR CANCELLED alert, or one is raised for a different reason,
|
|
hitting Refresh in the Combo tab establishes a connection, confirms the alert and shows
|
|
a notification for it in AAPS. This can safely be done, since those alerts are benign - an
|
|
appropriate TBR will be set again during the next loop iteration.
|
|
- For all other alerts raised by the pump: connecting to the pump will show the alert message in
|
|
the Combo tab, e.g. "State: E4: Occlusion" as well as showing a notification on the main screen.
|
|
An error will raise an urgent notification.
|
|
- After pairing, ruffy should not be used directly (AAPS will start in the background as needed),
|
|
since using ruffy at AAPS at the same time is not supported.
|
|
- If AAPS crashes (or is stopped from the debugger) while AAPS and the pump were comunicating (using
|
|
ruffy), it might be necessary to force close ruffy. Restarting AAPS will start ruffy again.
|
|
- Read the documentation on the wiki as well as the docs at https://openaps.org
|
|
- Don't press any buttons on the pump while AAPS communicates with the pump (Bluetooth logo is
|
|
shown on the pump).
|
|
- If the loop requests a running TBR to be cancelled the Combo will set a TBR of 90% or 110%
|
|
for 15 minutes instead. This is because cancelling a TBR causes an alert on the pump which
|
|
causes a lot of vibrations.
|
|
- Due to the bug (which causes the pump to become unreachable when reading history regularly),
|
|
a delivered bolus will NOT be added to treatments when the connection was lost during bolusing
|
|
or when a pump error occurs (e.g. occlusion). This will raise a message (and play an annoying
|
|
sound) and the user will have a create a bolus record via Careportal/NS manually. Currently
|
|
this requires an available NS installation and being online (for a short time). Hopefully
|
|
both of these issues can be resolved in future versions.
|
|
|
|
Known issues:
|
|
- Occasionally (every couple of days or so) AAPS might fail to automatically cancel
|
|
a TBR CANCELLED alert the user then needs to deal with (press the refresh button in AAPS
|
|
to transfer the warning to AAPS or confirm the alert on the pump).
|
|
- Similarly, the 'pump unreachable' bug may occur from time to time, which requires confirming
|
|
the alert on the pump to get the pump to accept connections again.
|
|
- Overall the integration seems rather robust, but there are limits to the way the
|
|
pump is controlled and how stable BT is, so there will be minor issues like the above
|
|
from time to time, though they're small compared to what works well.
|
|
- AAPS might be unresponsive for 10-30s or so when starting and calculating sensitivity.
|
|
AAPS might also be unresponsive when doing background work, e.g. after receiving a new
|
|
glucose reading.
|
|
- Since the pump's date & time can't be updated automatically at this time, daylight saving changes
|
|
will cause an alert asking to update the pump clock. A usable workaround should be to disable
|
|
the automatic update of the phone's clock for the night and then enabling it again in the morning
|
|
when also updating the pump.
|
|
|
|
Reporting bugs:
|
|
- Note the precise time the problem occurred and describe the circumstances and steps that caused
|
|
the problem
|
|
- Note the Build version (found in the About dialog in the app, when pressing the three dots in the
|
|
upper-right corner).
|
|
- Attach the app's log files, which can be found on the phone in
|
|
_/storage/emulated/0/Android/data/info.nightscout.androidaps/_
|
|
|
|
Components/Layers (developers):
|
|
- AndroidAPS
|
|
- ComboPlugin
|
|
- ruffy-spi (RuffyCommands interface to plug in lower layer)
|
|
- Scripting layer (ruffyscripter) / Command layer
|
|
- Driver layer (ruffy)
|