Readme/Testing updates.
This commit is contained in:
parent
28f31414ba
commit
e216bfcb12
2 changed files with 18 additions and 3 deletions
|
@ -23,7 +23,7 @@ Limitations:
|
|||
(e.g. bolusing from the pump and then immediately bolusing from AAPS).
|
||||
- If a TBR is set on the pump, AAPS will cancel it. This is because it's not possible to determine
|
||||
the start point of a TBR until it is finished or was cancelled at which point a record in the
|
||||
pumps history is created. Before that, there is none and it's simply not possible to determine
|
||||
pump's history is created. Before that, there is none and it's simply not possible to determine
|
||||
the TBRs influence on IOB. Set TBR using AAPS instead.
|
||||
- It's currently not possible to set the time and date on the pump (only reading the time - but
|
||||
not the date - is supported raises a warning in AAPS to update the pump clock manually).
|
||||
|
@ -32,8 +32,10 @@ Limitations:
|
|||
|
||||
Setup v2:
|
||||
- Configure pump using 360 config software.
|
||||
// TODO clarify with screenshot or at least better description of the on-screen items
|
||||
- Set/leave the menu configuration as "Standard"
|
||||
- Set/leave the menu configuration as "Standard", this will show only the supported
|
||||
menus/actions on the pump and hide those which are unsupported (extended/multiwave bolus,
|
||||
multiple basal rates), which cause the loop functionality to be disabled when used because
|
||||
it's not possible to run the loop in a safe manner when used.
|
||||
- Set maximum TBR to 500%
|
||||
- Disable end of TBR alert
|
||||
- Set low cartridge alarm to your licking
|
||||
|
|
|
@ -74,4 +74,17 @@
|
|||
must also raise an alert and disable loop
|
||||
- [ ] If a basal rate other than profile 1 is activated, this must also raise an alert and disable
|
||||
the loop for 6 h
|
||||
- [ ] Reading/setting basal profile
|
||||
- [ ] AAPS reads basal rate properly
|
||||
- [ ] AAPS doesn't touch pump basal if "Sync to profile" pref not set
|
||||
- [ ] AAPS updates basal rate if "Sync to profile" is enabled
|
||||
- [ ] Test profile with 115% (or something like that) change to ask the
|
||||
pump for basal rates like 0.812, which should then be set propely
|
||||
- [ ] Taking over alerts
|
||||
- [ ] If an error alert is active on the pump, pressing refresh shall display the error
|
||||
in the Combo tab but NOT confirm it. Easiest error to trigger: rewind piston
|
||||
and attempt to start the pump, this will trigger E11: Not primed.
|
||||
- [ ] Pressing refresh while a low cartridge or low battery alarm is active
|
||||
must confirm the alarm, indicate the new status in the Combo tab and
|
||||
show a notification on the overview screen
|
||||
|
||||
|
|
Loading…
Reference in a new issue