p5nbTgip0r 66bc927fea fix(medtronic): regression in cannula change detection
it seems an erroneous `lastPrimeRecordTime!=0L && ` was added during the kotlin refactorings of this:
59a3bf8883 (diff-b3c8866292a4a902f69ba7dc1b82892bae6fc93a2dce97749c34f1fcf9f1f751R400)

this commit just reverts that line, which restores the previously-working functionality of automatic cannula change detection in medtronic pumps
2022-05-22 09:56:59 -07:00
..
main fix(medtronic): regression in cannula change detection 2022-05-22 09:56:59 -07:00
test BgGraphBuilder -> kt 2022-04-23 18:34:04 +02:00