Johannes Mockenhaupt
467cf1e6ed
Pass pre-cmd pump state to command, log it.
2017-07-19 21:45:28 +02:00
Johannes Mockenhaupt
3b287df77d
Check active menu again when retrying inputs.
2017-07-18 21:20:40 +02:00
Johannes Mockenhaupt
ebe14fb0cf
BolusCommand: deal with blinking values.
...
How did this not trigger with all the boluses I've issued so far?
2017-07-18 21:20:40 +02:00
Johannes Mockenhaupt
7fe9ad7910
Add retries for inputing values.
...
Rarely there seem to be timing issues and e.g.
10 button down presses to go from 100% to 0% only goes down to 20%.
Retry two more times in that case, restarting the input process on the
active screen (bolus input, tbr percent/duration input).
2017-07-18 21:20:40 +02:00
Johannes Mockenhaupt
af4ec4d072
BolusCommand: fix contstraint check (doh!)
2017-07-18 21:20:40 +02:00
Johannes Mockenhaupt
44ae79bd50
Move command argument checking out of ctor into a dedicated method and check prior to execution.
2017-07-18 21:20:40 +02:00
Johannes Mockenhaupt
52b9621e29
Paranoia: check we're in the expected menu before each and every button press.
2017-07-18 21:20:40 +02:00
Johannes Mockenhaupt
97f9e99437
toString methods for Commands.
2017-07-18 20:55:47 +02:00
Johannes Mockenhaupt
1f5c03b64e
Cleanups.
2017-07-18 20:55:47 +02:00
Johannes Mockenhaupt
3280092566
Source ruffy scripter for the time being (already some fixes and tweaks in).
2017-07-18 20:55:44 +02:00