README updates.
This commit is contained in:
parent
4e526bba96
commit
a85da1d9ec
1 changed files with 12 additions and 12 deletions
|
@ -20,21 +20,20 @@ Limitations:
|
|||
- Setting a basal profile other than 1 on the pump, or delivering extended boluses or multiwave
|
||||
boluses from the pump interferes with TBRs and forces the loop into low-suspend only mode for 6 hours
|
||||
as the the loop can't run safely under those conditions.
|
||||
- It's currently not possible to set the time and date on the pump, however, this has no effect
|
||||
on AAPS and will not cause a problem if the pump's clock is not updated at the exact time
|
||||
daylight savings time starts or ends.
|
||||
- It's currently not possible to set the time and date on the pump, so daylight saving times
|
||||
changes have to be performed manually.
|
||||
- There's a bug in the pump's firmware that's triggered when "too much" communication happens
|
||||
with the pump. Specifically, this issue occurs when going from just issuing commands to the pump
|
||||
to reading the pumps data and history. For that reason, a minimal amount of data is read from
|
||||
the pump (no history).
|
||||
the pump.
|
||||
The bug might still rarely occur and causes the pump to not accept any connection
|
||||
unless a button is physically pressed on the pump.
|
||||
Therefore, the pump's reservoir level is not read and the pump status information uploaded to Nightscout
|
||||
shows fake numbers of 250 (above low threshold - which can be configured via the configuration
|
||||
tool), 50 (below low threshold) and 0 if the reservoir is empty.
|
||||
shows fake numbers of 150 (above low threshold - which can be configured via the configuration
|
||||
tool), 8 (below low threshold, triggers NS alarm since < 10) and 0 if the reservoir is empty.
|
||||
Furthermore, no history (from the My Data menu) is read unless absolutely required.
|
||||
Reading all pump data can be forced through the Combo tab (long press "TDDS"), to "import"
|
||||
events that happened solely on the pump, but that code has been tested less and may
|
||||
Reading all pump data can be forced through the Combo tab, to import
|
||||
events that happened solely on the pump, but that code has been tested less, might cause duplicates and may
|
||||
trigger the bug, so it's strongly recommended to stick to the usage scenario of controlling the
|
||||
pump solely through AAPS.
|
||||
Checking history, reservoir level etc on the pump causes no issues but should be avoided
|
||||
|
@ -67,7 +66,7 @@ Setup:
|
|||
to the pump.
|
||||
|
||||
Usage:
|
||||
- This is not a product, esp. in the beginning the user needs to monitor and understand the system,
|
||||
- Keep in mind that this is not a product, esp. in the beginning the user needs to monitor and understand the system,
|
||||
its limitations and how it can fail. It is strongly advised NOT to use this system when the person
|
||||
using is not able to fully understand the system.
|
||||
- This integration uses the same functionality which the meter provides that comes with the Combo.
|
||||
|
@ -121,17 +120,18 @@ Usage:
|
|||
both of these issues can be resolved in future versions.
|
||||
|
||||
Known issues:
|
||||
- On phones with low memory (or aggressive power saving setings, Android may kill
|
||||
- On phones with low memory (or aggressive power saving settings), Android may kill
|
||||
AAPS frequently (if the buttons on the overview screen aren't displayed when opening
|
||||
AAPS, the app was started again after Android killed it).
|
||||
This may trigger false 'pump unreachable alarms' on start.
|
||||
See the Combo tab's "last connection" field to check when the pump was last connected.
|
||||
This may drain the pump's battery quicker since on startup the basal profile is read
|
||||
from the pump. This may also increase the chance to hit the bug that makes the pump
|
||||
reject all incoming connections unless a button on the pump is pressed.
|
||||
reject all incoming connections until a button on the pump is pressed.
|
||||
- Occasionally (every couple of days or less) AAPS might fail to automatically cancel
|
||||
a TBR CANCELLED alert and needs to be dealt with (press the refresh button in AAPS
|
||||
to transfer the warning to AAPS or confirm the alert on the pump).
|
||||
to transfer the warning to AAPS or confirm the alert on the pump). Similarly, the
|
||||
'pump unreachable' bug may occur from time to time.
|
||||
- Overall the integration seems rather robust, but there are limits to the way the
|
||||
pump is controlled and how stable BT is, so there will be minor issues like the above
|
||||
from time to time, though they're small compared to what works well.
|
||||
|
|
Loading…