Fix TBR maximum for Animas Vibe

Animas actually supports 300% TBR, but uses a different way of
displaying it:

0% is the same as 100% in AAPS.
-100% (OFF) is 0% in AAPS
200% therefore is 300% in AAPS.

Maybe there should be a better indicator for this in AAPS, but for now
this fix gives better control of the TBR values, if using the open loop.
This commit is contained in:
Julius de Bruijn 2018-09-21 11:09:41 +02:00
parent 58b3f96c9e
commit 84e41122fc

View file

@ -56,7 +56,7 @@ public enum PumpType {
AnimasVibe("Animas Vibe", 0.05d, null, // AnimasBolus?
new DoseSettings(0.05d, 30, 12*60, 0.05d), //
PumpTempBasalType.Percent, //
new DoseSettings(10, 30, 24*60, 0d, 200d), PumpCapability.BasalRate_Duration30minAllowed, //
new DoseSettings(10, 30, 24*60, 0d, 300d), PumpCapability.BasalRate_Duration30minAllowed, //
0.025d, 5d, 0d, null, PumpCapability.VirtualPumpCapabilities), //
AnimasPing("Animas Ping", AnimasVibe),