Updating Testing-Combo.md to reflect removed history functionality.
This commit is contained in:
parent
a38afe7a2a
commit
6e5f848c6b
|
@ -3,25 +3,25 @@
|
|||
- [ ] Bolusing
|
||||
- [ ] Cancelling bolus at various stages and checking error message (if any) and check
|
||||
no bolus or a partial bolus was delivered
|
||||
- [ ] Cancelling a bolus 0.2 U before its finished (cancelling requires pressing up/down
|
||||
- [ ] Cancelling a bolus 0.2 U before it's finished (cancelling requires pressing up/down
|
||||
3s, so the cancellation will be too late, but the code should handle it and report
|
||||
the full bolus being delivered
|
||||
- [ ] Low cartridge alarm during bolus
|
||||
- [ ] must be confirmed by AAPS
|
||||
- [ ] alarm must be confirmed by AAPS
|
||||
- [ ] bolus must have been fully delivered by pump
|
||||
- [ ] bolus must have been added to DB
|
||||
- [ ] the confirmed pump warning must be raised as a notification in AAPS
|
||||
- [ ] notification must also appear on smartwatch
|
||||
- [ ] An error during bolus must yield an error in AAPS
|
||||
- [ ] An error during bolus must yield a notification on a smartwatch
|
||||
- [ ] Bolusing e.g. 4 U if reservoir has only 2 U left must be rejected
|
||||
- [ ] Bolusing e.g. 4 U if reservoir has only 2 U must yield a usable error
|
||||
- [ ] Pressing a button on the pump before bolus delivery started must be handled gracefully
|
||||
- [ ] Same as above, but moving pump out of range
|
||||
- [ ] Pressing a button on the pump after bolus delivery has started must be handled
|
||||
gracefully (bolus finished and AAPS must correctly add it the the DB)
|
||||
- [ ] Same as above, but moving pump out of range
|
||||
- [ ] Test the highest bolus you'd ever give yourself (AAPS has a configurable limit and the pump
|
||||
has a limit which can be configured with the Config SW)
|
||||
has a limit which can be configured with the Config SW), no timeout or other issues must show
|
||||
- [ ] Test a bolus exceeding the pumps max bolus; should show a generic error message
|
||||
- [ ] BT disconnect issues
|
||||
- [ ] Moving pump out of reach when setting TBR causes "TBR cancelled" alarm on pump.
|
||||
|
@ -30,9 +30,8 @@
|
|||
- [ ] Same as above while bolusing must report an error and NOT retry the command
|
||||
- [ ] AAPS start
|
||||
- [ ] Starting AAPS without a reachable pump must show something sensible in the Combo tab
|
||||
(that is not hang indefinitely with "initializing" activity)
|
||||
- [ ] Starting AAPS without a reachable pump must trigger "pump unrechable" alert after
|
||||
5 to 10 min.
|
||||
(not hanging indefinitely with "initializing" activity)
|
||||
- [ ] Starting AAPS without a reachable pump must trigger "pump unrechable" alert after the configured threshold
|
||||
- [ ] If the pump's profile doesn't match AAPS', the pump must be updated when AAPS starts
|
||||
- [ ] Doing a profile change (to shift time or increase/decrease insulin), the pump's basal profile must be updated
|
||||
- [ ] If a profile change has a duration, the pump's basal profile must be set to the original value again at the end
|
||||
|
@ -43,12 +42,14 @@
|
|||
- [ ] Boluses
|
||||
- [ ] TBR
|
||||
- [ ] Alerts
|
||||
- [ ] TDDs
|
||||
- [ ] Disconnected pump (pump unreachable)
|
||||
- [ ] ...
|
||||
- [ ] With local alerts enabled for 'pump unreachable', an alert must be triggered within 5 minutes
|
||||
after the configured threshold. (Don't set the threshold too low, e.g. 10 minutes, since
|
||||
there might be no need to set a TBR within such a short time).
|
||||
- [ ] Refilling cartridge
|
||||
- [ ] If TBR was cancelled by refilling, AAPS must detect this and update the TBR treatment
|
||||
accordingly
|
||||
- [ ] If TBR was cancelled by refilling, AAPS must detect this and create a TBR record in AAPS
|
||||
based on what the pump displays (not the full TBR duration, but what is displayed as remaining
|
||||
on the main screen.
|
||||
- [ ] Stress testing
|
||||
- PersistentNotification plugin disabled
|
||||
- Lots of comms running, like Wifi, GSM, BT audio
|
||||
|
@ -58,22 +59,24 @@
|
|||
- [ ] With the pump powered off or out of reach, the 'pump unreachable alert' must still
|
||||
trigger
|
||||
- [ ] Combo tab
|
||||
- [ ] Check displayed data (state, battery, reservoir, last bolus, temp basal) is the same
|
||||
- [ ] Check displayed data (state, battery, reservoir, bolus, temp basal) is the same
|
||||
as on the pump
|
||||
- [ ] Unsafe usage
|
||||
- [ ] An extended or multiwave bolus given within the last six hour must raise an alert and
|
||||
closed loop functionality (Loop plugin shows "disabled by constrains" when "execute" is pressed)
|
||||
restrict the loop functionality to low-suspend only (setting maxIOB to zero)
|
||||
- [ ] Closed loop functionality must resume 6 h after the last ext/multiwave bolus
|
||||
- [ ] An active ext/multiwave bolus (a history record is created only after the bolus completed)
|
||||
must also raise an alert and disable loop
|
||||
must also raise an alert and restrict the loop
|
||||
- [ ] If a basal rate other than profile 1 is activated, this must also raise an alert and disable
|
||||
the loop for 6 h
|
||||
the restrict the loop
|
||||
- [ ] Reading/setting basal profile
|
||||
- [ ] AAPS reads basal rate properly
|
||||
- [ ] AAPS doesn't touch pump basal if "Sync to profile" pref not set
|
||||
- [ ] AAPS updates basal rate if "Sync to profile" is enabled
|
||||
- [ ] Test profile with 115% (or something like that) change to ask the
|
||||
pump for basal rates like 0.812, which should then be set propely
|
||||
- [ ] Updating the profile extensively (200%, shifting time) takes up to 6 minutes, but
|
||||
should complete without timeout.
|
||||
- [ ] Taking over alerts
|
||||
- [ ] If an error alert is active on the pump, pressing refresh shall display the error
|
||||
in the Combo tab but NOT confirm it. Easiest error to trigger: rewind piston
|
||||
|
|