Testing-Combo: add entry on experimenting with menu timeouts.

This commit is contained in:
Johannes Mockenhaupt 2017-12-08 23:49:33 +01:00
parent 2c8c25c082
commit 6db58a4e93
No known key found for this signature in database
GPG key ID: 9E1EA6AF7BBBB0D1

View file

@ -27,6 +27,10 @@
- [ ] Moving pump out of reach when setting TBR causes "TBR cancelled" alarm on pump.
When putting pump close to phone, AAPS must confirm the alert and successfully
retry setting TBR (reconnects are a best-effort kind of a thing, so this might not always work)
- [ ] When a disconnect occurs, the pump's screen shows the error and the pump only accepts a connection
again when a timeout has occurred. A recovery should be quicker if that timeout is decreased.
It might be interesting to experiment with the Config software to set lower menu or display timeouts
(or whatever they're called ...) to improve recovery speed.
- [ ] Same as above while bolusing must report an error and NOT retry the command
- [ ] AAPS start
- [ ] Starting AAPS without a reachable pump must show something sensible in the Combo tab