@ -450,6 +450,26 @@ public class RuffyScripter {
log.debug("Releasing back key");
}
public void pressKeyMs(final byte key, long ms) {
long stepMs = 100;
try {
log.debug("Scroll: Pressing key for " + ms + " ms with step " + stepMs + " ms");
ruffyService.rtSendKey(key, true);
ruffyService.rtSendKey(key, false);
while (ms > stepMs) {
SystemClock.sleep(stepMs);
ruffyService.rtSendKey(key, false);
ms -= stepMs;
}
SystemClock.sleep(ms);
ruffyService.rtSendKey(Key.NO_KEY, true);
log.debug("Releasing key");
} catch (Exception e) {
throw new CommandException().exception(e).message("Error while pressing buttons");
}
}
public boolean waitForScreenUpdate(long timeout) {
synchronized (screenlock) {
try {

View file

@ -84,7 +84,7 @@ public class BolusCommand implements Command {
progressReportCallback.report(DELIVERING, 0, 0);
Double bolusRemaining = (Double) scripter.currentMenu.getAttribute(MenuAttribute.BOLUS_REMAINING);
double lastBolusReported = 0;
List<String> alarmsRaised = new ArrayList<>();
boolean lowCartdrigeAlarmTriggered = false;
// wait for bolus delivery to complete; the remaining units to deliver are counted
// down and are displayed on the main menu.
// TODO extract into method
@ -94,20 +94,19 @@ public class BolusCommand implements Command {
while (bolusRemaining != null) {
if (cancelRequested) {
progressReportCallback.report(STOPPING, 0, 0);
// TODO just press up 3s in a separated thread and let this loop run
// and at the end handle the outcome and returned raise alarms, whether cancel was reuqested etc
// TODO press up 3s, deal with bolus cancelled error, retrieved amount actually delivered from history and return it
// since the cancellation takes three seconds some insulin will have definitely been delivered (delivery speed is roughly 0.1U/s),
// but the pump may also finish delivering the bolus while we try to cancel it
// so, press a button, keep it press and deal with three outcomes:
// * delivery finished (no more remaining bolus displayed)
// * bolus was cancelled (warning raised)
// * any other error (low cartridge, occlusion, both will also trigger 'bolus cancelled' errors)
// cancelBolusInDelivery()
// TODO new thread to press button and then deal with outcomes below, since all errors can occur at all time, pressing
// abort just forces an error (if keyrpess is in time)
scripter.pressKeyMs(RuffyScripter.Key.UP, 3000);
progressReportCallback.report(STOPPED, 0, 0);
// if the bolus finished while we attempted to cancel it, there'll be no alarm
long timeout = System.currentTimeMillis() + 2000;
while (scripter.currentMenu.getType() != MenuType.WARNING_OR_ERROR && System.currentTimeMillis() < timeout) {
SystemClock.sleep(10);
}
while (scripter.currentMenu.getType() == MenuType.WARNING_OR_ERROR) {
// TODO make this cleaner, extract method, needed below too
scripter.pressCheckKey();
SystemClock.sleep(200);
}
break;
}
if (lastBolusReported != bolusRemaining) {
log.debug("Delivering bolus, remaining: " + bolusRemaining);
@ -115,23 +114,12 @@ public class BolusCommand implements Command {
progressReportCallback.report(DELIVERING, percentDelivered, bolus - bolusRemaining);
lastBolusReported = bolusRemaining;
}
// TODO deal with alarms that can arise; an oclussion with raise an oclussion alert as well as a bolus cancelled alert
// occlusion cancels the bolus -> abort routine to report back delivered bolus;
// low cartridge alert lets bolus run out
// also, any other error or warning can occur and we should return in a controlled fashion -
// communicating back what was actually delivered.
// generally: cancel an alert on the pump and raise the error in AAPS?
// letting the alert go off disrupts comms if the user interacts with the pump,
// then we need to schedule a history read in the near future, let thee user know
// the data will be out of sync for a bit.
// how does the dana handle pump errors? has no vibration, but sound i guess
// should this be configurabe? initially?
if (scripter.currentMenu.getType() == MenuType.WARNING_OR_ERROR) {
String message = (String) scripter.currentMenu.getAttribute(MenuAttribute.MESSAGE);
if (message.equals("LOW CARTRIDGE")) {
alarmsRaised.add(message);
// confirm, note alert was raised and continue bolusing)
lowCartdrigeAlarmTriggered = true;
confirmAlert("LOW CARTRIDGE", 2000);
} else {
// any other alert
break;
@ -141,7 +129,7 @@ public class BolusCommand implements Command {
bolusRemaining = (Double) scripter.currentMenu.getAttribute(MenuAttribute.BOLUS_REMAINING);
}
// wait up to 2s for any possible warning to be raised
// wait up to 2s for any possible warning to be raised, if not raised already
long minWait = System.currentTimeMillis() + 2 * 1000;
while (scripter.currentMenu.getType() != MenuType.WARNING_OR_ERROR || System.currentTimeMillis() < minWait) {
SystemClock.sleep(50);
@ -177,6 +165,8 @@ public class BolusCommand implements Command {
.message("Bolus was delivered, but unable to confirm it with history record");
}
// TODO check date so we don't pick a false record if the previous bolus had the same amount;
// also, report back partial bolus. Just call ReadHsstory(timestamp, boluses=true) cmd ...
double lastBolusInHistory = (double) scripter.currentMenu.getAttribute(MenuAttribute.BOLUS);
if (Math.abs(bolus - lastBolusInHistory) > 0.05) {
throw new CommandException().success(false).enacted(true)