combov2: Consider pump as uninitialized for a while after an error occurred

Signed-off-by: Carlos Rafael Giani <crg7475@mailbox.org>
This commit is contained in:
Carlos Rafael Giani 2022-12-23 17:02:33 +01:00
parent b39f5a3672
commit 6b989ec7aa
3 changed files with 57 additions and 2 deletions

View file

@ -41,6 +41,7 @@ class ComboV2Fragment : DaggerFragment() {
binding.combov2RefreshButton.setOnClickListener {
binding.combov2RefreshButton.isEnabled = false
combov2Plugin.clearPumpErrorObservedFlag()
commandQueue.readStatus(rh.gs(info.nightscout.core.ui.R.string.user_request), null)
}

View file

@ -95,6 +95,8 @@ import info.nightscout.comboctl.base.Tbr as ComboCtlTbr
import info.nightscout.comboctl.main.Pump as ComboCtlPump
import info.nightscout.comboctl.main.PumpManager as ComboCtlPumpManager
internal const val PUMP_ERROR_TIMEOUT_INTERVAL_MSECS = 1000L * 60 * 5
@Singleton
class ComboV2Plugin @Inject constructor (
injector: HasAndroidInjector,
@ -161,6 +163,13 @@ class ComboV2Plugin @Inject constructor (
private var lastConnectionTimestamp = 0L
private var lastComboAlert: AlertScreenContent? = null
// States for when the pump reports an error. We then want isInitialized()
// to return false until either the user presses the Refresh button or the
// pumpErrorTimeoutJob expires. That way, the loop won't run until then,
// giving the user a chance to handle the error.
private var pumpErrorObserved = false
private var pumpErrorTimeoutJob: Job? = null
// Set to true if a disconnect request came in while the driver
// was in the Connecting, CheckingPump, or ExecutingCommand
// state (in other words, while isBusy() was returning true).
@ -183,7 +192,7 @@ class ComboV2Plugin @Inject constructor (
private var activeBasalProfile: BasalProfile? = null
// This is used for checking that the correct basal profile is
// active in the Combo. If not, loop invocation is disallowed.
// This is _not_ reset by disconect(). That's on purpose; it
// This is _not_ reset by disconnect(). That's on purpose; it
// is read by isLoopInvocationAllowed(), which is called even
// if the pump is not connected.
private var lastActiveBasalProfileNumber: Int? = null
@ -381,7 +390,7 @@ class ComboV2Plugin @Inject constructor (
}
override fun isInitialized(): Boolean =
isPaired() && (driverStateFlow.value != DriverState.NotInitialized)
isPaired() && (driverStateFlow.value != DriverState.NotInitialized) && !pumpErrorObserved
override fun isSuspended(): Boolean =
when (driverStateUIFlow.value) {
@ -432,6 +441,16 @@ class ComboV2Plugin @Inject constructor (
return
}
if (pumpErrorObserved) {
aapsLogger.debug(LTag.PUMP, "Aborting connect attempt since the pumpErrorObserved flag is set")
uiInteraction.addNotification(
Notification.COMBO_PUMP_ALARM,
text = rh.gs(R.string.combov2_cannot_connect_pump_error_observed),
level = Notification.NORMAL
)
return
}
when (driverStateFlow.value) {
DriverState.Connecting,
DriverState.CheckingPump,
@ -1398,6 +1417,14 @@ class ComboV2Plugin @Inject constructor (
commandQueue.readStatus(reason, null)
}
fun clearPumpErrorObservedFlag() {
stopPumpErrorTimeout()
if (pumpErrorObserved) {
aapsLogger.info(LTag.PUMP, "Clearing pumpErrorObserved flag")
pumpErrorObserved = false
}
}
/*** Loop constraints ***/
// These restrict the function of the loop in case of an event
// that makes running a loop too risky, for example because something
@ -1550,6 +1577,8 @@ class ComboV2Plugin @Inject constructor (
_serialNumberUIFlow.value = ""
_bluetoothAddressUIFlow.value = ""
clearPumpErrorObservedFlag()
// The unpairing variable is set to false in
// the PumpManager onPumpUnpaired callback.
}
@ -1748,6 +1777,23 @@ class ComboV2Plugin @Inject constructor (
}
}
private fun startPumpErrorTimeout() {
if (pumpErrorTimeoutJob != null)
return
pumpErrorTimeoutJob = pumpCoroutineScope.launch {
delay(PUMP_ERROR_TIMEOUT_INTERVAL_MSECS)
aapsLogger.info(LTag.PUMP, "Clearing pumpErrorObserved flag after timeout was reached")
pumpErrorObserved = false
commandQueue.readStatus(rh.gs(R.string.combov2_refresh_pump_status_after_error), null)
}
}
private fun stopPumpErrorTimeout() {
pumpErrorTimeoutJob?.cancel()
pumpErrorTimeoutJob = null
}
private fun updateBaseBasalRateUI() {
val currentHour = DateTime().hourOfDay().get()
// This sets value to null if no profile is set,
@ -2142,6 +2188,12 @@ class ComboV2Plugin @Inject constructor (
}
private fun notifyAboutComboAlert(alert: AlertScreenContent) {
if (alert is AlertScreenContent.Error) {
aapsLogger.info(LTag.PUMP, "Error screen observed - setting pumpErrorObserved flag")
pumpErrorObserved = true
startPumpErrorTimeout()
}
uiInteraction.addNotification(
Notification.COMBO_PUMP_ALARM,
text = "${rh.gs(R.string.combov2_combo_alert)}: ${getAlertDescription(alert)}",

View file

@ -135,4 +135,6 @@ buttons at the same time to cancel pairing)\n
<string name="combov2_datetime_changed">Date and/or time changed</string>
<string name="combov2_dst_started">Daylight savings time (DST) started</string>
<string name="combov2_dst_ended">Daylight savings time (DST) ended</string>
<string name="combov2_cannot_connect_pump_error_observed">Cannot connect to pump because the pump reported an error. User must handle the error and then either wait 5 minutes or press the Refresh button in the driver tab.</string>
<string name="combov2_refresh_pump_status_after_error">Refreshing pump status after the pump reported an error</string>
</resources>