Clean up Testing-Combo.md a bit.
This commit is contained in:
parent
a85da1d9ec
commit
3ac63a980c
1 changed files with 33 additions and 35 deletions
|
@ -1,43 +1,14 @@
|
|||
- [ ] Pairing
|
||||
- [ ] Pairing works with `combo-scripter-v2` branch
|
||||
- [ ] Bolusing
|
||||
- [ ] Cancelling bolus at various stages and checking error message (if any) and check
|
||||
no bolus or a partial bolus was delivered
|
||||
- [ ] Cancelling bolus at various stages shall not yield an error but cancel the bolus. If
|
||||
cancelled before delivery started, no treatment must have been added, if cancel after delivery
|
||||
started, the partially delivered bolus must have been added to treatments
|
||||
- [ ] Enter a bolus of 2 U and press cancel when delivery is at 1.7 (cancelling requires AAAPS
|
||||
to press the up button for 3 seconds, so the cancellation attempt will not succeed because delivery
|
||||
ends before those 3 seconds are elapsed). The code should handle this without giving an
|
||||
error and add the full bolus to treatments.
|
||||
- [ ] Low cartridge alarm during bolus
|
||||
- [ ] alarm must be confirmed by AAPS
|
||||
- [ ] bolus must have been fully delivered by pump
|
||||
- [ ] bolus must have been added to DB
|
||||
- [ ] the confirmed pump warning must be raised as a notification in AAPS
|
||||
(or as android notification on watch/smartphone if setting "use system notifications ..." is enabled
|
||||
- [ ] An error during bolus must yield an error in AAPS
|
||||
- [ ] An error during bolus must yield a notification on a smartwatch
|
||||
- [ ] Test bolusing a bolus bigger than what's left in the reservoir. A message to check what
|
||||
was actually delivered must appear (this is a corner-case where we practically can't
|
||||
check what was actually delivered).
|
||||
- [ ] Pressing a button on the pump before bolus delivery started must be handled gracefully
|
||||
- [ ] Same as above, but moving pump out of range
|
||||
- [ ] Pressing a button on the pump after bolus delivery has started must raise a warning
|
||||
to check what was actually delivered and manually update treatments in AAPS.
|
||||
- [ ] Same as above, but moving pump out of range
|
||||
- [ ] Test the highest bolus you'd ever give yourself (AAPS has a configurable limit and the pump
|
||||
has a limit which can be configured with the Config SW), no timeout or other issues must show
|
||||
- [ ] Test a bolus exceeding the pumps max bolus; should show a generic error message
|
||||
- [ ] BT disconnect issues
|
||||
- [ ] Moving pump out of reach when setting TBR causes "TBR cancelled" alarm on pump.
|
||||
When putting pump close to phone, AAPS must confirm the alert and successfully
|
||||
retry setting TBR (reconnects are a best-effort kind of a thing, so this might not always work)
|
||||
- [ ] When a disconnect occurs, the pump's screen shows the error and the pump only accepts a connection
|
||||
again when a timeout has occurred. A recovery should be quicker if that timeout is decreased.
|
||||
It might be interesting to experiment with the Config software to set lower menu or display timeouts
|
||||
(or whatever they're called ...) to improve recovery speed.
|
||||
- [ ] Same as above while bolusing must report an error and NOT retry the command
|
||||
- [ ] Recovery from connection issues during bolusing
|
||||
- [ ] Bolusing still works => No error dialog, record is added to treatments
|
||||
- [ ] Cancelling the bolus still works (while bolus is in progress)
|
||||
- [ ] Pressing a button on the pump during delivery => Progress dialog freezes, then states that recovery
|
||||
is in process and then closes; no error dialog, record correctly added to treatments
|
||||
- [ ] Breaking the connection e.g. by moving the pump away from phone for up to a minute => same as above
|
||||
|
@ -45,11 +16,38 @@
|
|||
- [ ] Starting a bolus bigger than what's left in the reservoir => Error dialog and a record in treatments with the partially delivered bolus
|
||||
- [ ] When the connection breaks during bolusing, pressing the cancel button should not interfere with recovery and
|
||||
the delivered bolus should be added to treatments
|
||||
- [ ] Low cartridge alarm during bolus
|
||||
- [ ] alarm must be confirmed by AAPS
|
||||
- [ ] bolus must have been fully delivered by pump
|
||||
- [ ] bolus must have been added to DB
|
||||
- [ ] the confirmed pump warning must be raised as a notification in AAPS
|
||||
(or as android notification on watch/smartphone if setting "use system notifications ..." is enabled
|
||||
- [ ] Pressing a button on the pump, or moving the pump away from the phone to break connection
|
||||
must confirm the pump alert and recover to finish the bolusing.
|
||||
- [ ] If recovery fails, an error popup must be displayed
|
||||
- [ ] Test bolusing a bolus bigger than what's left in the reservoir. A message to check what
|
||||
was actually delivered must appear (this is a corner-case where we practically can't
|
||||
check what was actually delivered).
|
||||
- [ ] Pressing a button on the pump before bolus delivery started must be handled gracefully
|
||||
- [ ] Same as above, but moving pump out of range
|
||||
- [ ] Pressing a button on the pump after bolus delivery has started will freeze the progress bar,
|
||||
initiate recovery and add the delivered bolus the treatments.
|
||||
- [ ] Same as above, but moving pump out of range
|
||||
- [ ] Test the highest bolus you'd ever give yourself (AAPS has a configurable limit and the pump
|
||||
has a limit which can be configured with the Config SW), no timeout or other issues must show
|
||||
- [ ] BT disconnect issues
|
||||
- [ ] Moving pump out of reach when setting TBR causes "TBR cancelled" alarm on pump.
|
||||
When putting pump close to phone, AAPS must confirm the alert and successfully
|
||||
retry setting TBR (reconnects are a best-effort kind of a thing, so this might not always work)
|
||||
- [ ] When a disconnect occurs, the pump's screen shows the error and the pump only accepts a connection
|
||||
again when the pump's menu(?) timeout has occurred. A recovery should be quicker if that timeout is decreased.
|
||||
It might be interesting to experiment with the Config software to set lower menu or display timeouts
|
||||
(or whatever they're called ...) to improve recovery speed.
|
||||
- [ ] AAPS start
|
||||
- [ ] Starting AAPS without a reachable pump must show something sensible in the Combo tab
|
||||
(not hanging indefinitely with "initializing" activity)
|
||||
- [ ] Starting AAPS without a reachable pump must trigger "pump unrechable" alert after the configured threshold
|
||||
- [ ] If the pump's profile doesn't match AAPS', the pump must be updated when AAPS starts
|
||||
- [ ] If the pump's basal profile doesn't match AAPS', the pump must be updated when AAPS starts
|
||||
- [ ] Read history using Smartpix and compare with AAPS' DB (treatment tab)
|
||||
Esp. those times we communication was interrupted, boluses were cancelled, ...
|
||||
- [ ] Boluses
|
||||
|
@ -58,7 +56,8 @@
|
|||
- [ ] Disconnected pump (pump unreachable)
|
||||
- [ ] With local alerts enabled for 'pump unreachable', an alert must be triggered within 5 minutes
|
||||
after the configured threshold. (Don't set the threshold too low, e.g. 10 minutes, since
|
||||
there might be no need to set a TBR within such a short time).
|
||||
there might be no need to set a TBR within such a short time, but since there was no pump connection
|
||||
within that time, the alarm would be triggered).
|
||||
- [ ] Refilling cartridge
|
||||
- [ ] If TBR was cancelled by refilling, AAPS must detect this and create a TBR record in AAPS
|
||||
based on what the pump displays (not the full TBR duration, but what is displayed as remaining
|
||||
|
@ -103,4 +102,3 @@
|
|||
- [ ] Misc
|
||||
- [ ] Pump state is correctly uploaded to Nightscout (note that reservoir level are fake numbers representing
|
||||
norma/low/empty).
|
||||
|
||||
|
|
Loading…
Reference in a new issue