Clean up Testing-Combo.md a bit.

This commit is contained in:
Johannes Mockenhaupt 2017-12-17 16:56:52 +01:00
parent a85da1d9ec
commit 3ac63a980c
No known key found for this signature in database
GPG key ID: 9E1EA6AF7BBBB0D1

View file

@ -1,43 +1,14 @@
- [ ] Pairing - [ ] Pairing
- [ ] Pairing works with `combo-scripter-v2` branch - [ ] Pairing works with `combo-scripter-v2` branch
- [ ] Bolusing - [ ] Bolusing
- [ ] Cancelling bolus at various stages and checking error message (if any) and check - [ ] Cancelling bolus at various stages shall not yield an error but cancel the bolus. If
no bolus or a partial bolus was delivered cancelled before delivery started, no treatment must have been added, if cancel after delivery
started, the partially delivered bolus must have been added to treatments
- [ ] Enter a bolus of 2 U and press cancel when delivery is at 1.7 (cancelling requires AAAPS - [ ] Enter a bolus of 2 U and press cancel when delivery is at 1.7 (cancelling requires AAAPS
to press the up button for 3 seconds, so the cancellation attempt will not succeed because delivery to press the up button for 3 seconds, so the cancellation attempt will not succeed because delivery
ends before those 3 seconds are elapsed). The code should handle this without giving an ends before those 3 seconds are elapsed). The code should handle this without giving an
error and add the full bolus to treatments. error and add the full bolus to treatments.
- [ ] Low cartridge alarm during bolus
- [ ] alarm must be confirmed by AAPS
- [ ] bolus must have been fully delivered by pump
- [ ] bolus must have been added to DB
- [ ] the confirmed pump warning must be raised as a notification in AAPS
(or as android notification on watch/smartphone if setting "use system notifications ..." is enabled
- [ ] An error during bolus must yield an error in AAPS
- [ ] An error during bolus must yield a notification on a smartwatch
- [ ] Test bolusing a bolus bigger than what's left in the reservoir. A message to check what
was actually delivered must appear (this is a corner-case where we practically can't
check what was actually delivered).
- [ ] Pressing a button on the pump before bolus delivery started must be handled gracefully
- [ ] Same as above, but moving pump out of range
- [ ] Pressing a button on the pump after bolus delivery has started must raise a warning
to check what was actually delivered and manually update treatments in AAPS.
- [ ] Same as above, but moving pump out of range
- [ ] Test the highest bolus you'd ever give yourself (AAPS has a configurable limit and the pump
has a limit which can be configured with the Config SW), no timeout or other issues must show
- [ ] Test a bolus exceeding the pumps max bolus; should show a generic error message
- [ ] BT disconnect issues
- [ ] Moving pump out of reach when setting TBR causes "TBR cancelled" alarm on pump.
When putting pump close to phone, AAPS must confirm the alert and successfully
retry setting TBR (reconnects are a best-effort kind of a thing, so this might not always work)
- [ ] When a disconnect occurs, the pump's screen shows the error and the pump only accepts a connection
again when a timeout has occurred. A recovery should be quicker if that timeout is decreased.
It might be interesting to experiment with the Config software to set lower menu or display timeouts
(or whatever they're called ...) to improve recovery speed.
- [ ] Same as above while bolusing must report an error and NOT retry the command
- [ ] Recovery from connection issues during bolusing - [ ] Recovery from connection issues during bolusing
- [ ] Bolusing still works => No error dialog, record is added to treatments
- [ ] Cancelling the bolus still works (while bolus is in progress)
- [ ] Pressing a button on the pump during delivery => Progress dialog freezes, then states that recovery - [ ] Pressing a button on the pump during delivery => Progress dialog freezes, then states that recovery
is in process and then closes; no error dialog, record correctly added to treatments is in process and then closes; no error dialog, record correctly added to treatments
- [ ] Breaking the connection e.g. by moving the pump away from phone for up to a minute => same as above - [ ] Breaking the connection e.g. by moving the pump away from phone for up to a minute => same as above
@ -45,11 +16,38 @@
- [ ] Starting a bolus bigger than what's left in the reservoir => Error dialog and a record in treatments with the partially delivered bolus - [ ] Starting a bolus bigger than what's left in the reservoir => Error dialog and a record in treatments with the partially delivered bolus
- [ ] When the connection breaks during bolusing, pressing the cancel button should not interfere with recovery and - [ ] When the connection breaks during bolusing, pressing the cancel button should not interfere with recovery and
the delivered bolus should be added to treatments the delivered bolus should be added to treatments
- [ ] Low cartridge alarm during bolus
- [ ] alarm must be confirmed by AAPS
- [ ] bolus must have been fully delivered by pump
- [ ] bolus must have been added to DB
- [ ] the confirmed pump warning must be raised as a notification in AAPS
(or as android notification on watch/smartphone if setting "use system notifications ..." is enabled
- [ ] Pressing a button on the pump, or moving the pump away from the phone to break connection
must confirm the pump alert and recover to finish the bolusing.
- [ ] If recovery fails, an error popup must be displayed
- [ ] Test bolusing a bolus bigger than what's left in the reservoir. A message to check what
was actually delivered must appear (this is a corner-case where we practically can't
check what was actually delivered).
- [ ] Pressing a button on the pump before bolus delivery started must be handled gracefully
- [ ] Same as above, but moving pump out of range
- [ ] Pressing a button on the pump after bolus delivery has started will freeze the progress bar,
initiate recovery and add the delivered bolus the treatments.
- [ ] Same as above, but moving pump out of range
- [ ] Test the highest bolus you'd ever give yourself (AAPS has a configurable limit and the pump
has a limit which can be configured with the Config SW), no timeout or other issues must show
- [ ] BT disconnect issues
- [ ] Moving pump out of reach when setting TBR causes "TBR cancelled" alarm on pump.
When putting pump close to phone, AAPS must confirm the alert and successfully
retry setting TBR (reconnects are a best-effort kind of a thing, so this might not always work)
- [ ] When a disconnect occurs, the pump's screen shows the error and the pump only accepts a connection
again when the pump's menu(?) timeout has occurred. A recovery should be quicker if that timeout is decreased.
It might be interesting to experiment with the Config software to set lower menu or display timeouts
(or whatever they're called ...) to improve recovery speed.
- [ ] AAPS start - [ ] AAPS start
- [ ] Starting AAPS without a reachable pump must show something sensible in the Combo tab - [ ] Starting AAPS without a reachable pump must show something sensible in the Combo tab
(not hanging indefinitely with "initializing" activity) (not hanging indefinitely with "initializing" activity)
- [ ] Starting AAPS without a reachable pump must trigger "pump unrechable" alert after the configured threshold - [ ] Starting AAPS without a reachable pump must trigger "pump unrechable" alert after the configured threshold
- [ ] If the pump's profile doesn't match AAPS', the pump must be updated when AAPS starts - [ ] If the pump's basal profile doesn't match AAPS', the pump must be updated when AAPS starts
- [ ] Read history using Smartpix and compare with AAPS' DB (treatment tab) - [ ] Read history using Smartpix and compare with AAPS' DB (treatment tab)
Esp. those times we communication was interrupted, boluses were cancelled, ... Esp. those times we communication was interrupted, boluses were cancelled, ...
- [ ] Boluses - [ ] Boluses
@ -58,7 +56,8 @@
- [ ] Disconnected pump (pump unreachable) - [ ] Disconnected pump (pump unreachable)
- [ ] With local alerts enabled for 'pump unreachable', an alert must be triggered within 5 minutes - [ ] With local alerts enabled for 'pump unreachable', an alert must be triggered within 5 minutes
after the configured threshold. (Don't set the threshold too low, e.g. 10 minutes, since after the configured threshold. (Don't set the threshold too low, e.g. 10 minutes, since
there might be no need to set a TBR within such a short time). there might be no need to set a TBR within such a short time, but since there was no pump connection
within that time, the alarm would be triggered).
- [ ] Refilling cartridge - [ ] Refilling cartridge
- [ ] If TBR was cancelled by refilling, AAPS must detect this and create a TBR record in AAPS - [ ] If TBR was cancelled by refilling, AAPS must detect this and create a TBR record in AAPS
based on what the pump displays (not the full TBR duration, but what is displayed as remaining based on what the pump displays (not the full TBR duration, but what is displayed as remaining
@ -103,4 +102,3 @@
- [ ] Misc - [ ] Misc
- [ ] Pump state is correctly uploaded to Nightscout (note that reservoir level are fake numbers representing - [ ] Pump state is correctly uploaded to Nightscout (note that reservoir level are fake numbers representing
norma/low/empty). norma/low/empty).