Document soft TBR cancel, other docs updates.

This commit is contained in:
Johannes Mockenhaupt 2017-11-25 10:51:54 +01:00
parent e13f29e615
commit 32112597bc
No known key found for this signature in database
GPG key ID: 9E1EA6AF7BBBB0D1

View file

@ -41,12 +41,17 @@ Setup v2:
- Disable end of TBR alert - Disable end of TBR alert
- Set low cartridge alarm to your licking - Set low cartridge alarm to your licking
- Enable keylock (can also be set on the pump directly, see usage section on reasoning) - Enable keylock (can also be set on the pump directly, see usage section on reasoning)
- Get Android Studio 3
- Get ruffy from https://github.com/jotomo/ruffy (branch `combo-scripter-v2`) - Get ruffy from https://github.com/jotomo/ruffy (branch `combo-scripter-v2`)
- Pair the pump, if it doesn't work, switch to the `pairing` branch, pair, - Pair the pump, if it doesn't work, switch to the `pairing` branch, pair,
then switch back the original branch. Note that AndroidAPS must not try to then switch back the original branch. If the pump is already paired and
talk to the pump during the pairing process TODO setting MDI sufficient? can be controlled via ruffy, installing the above version is sufficient.
- Get AndroidAPS from https://gitlab.com/jotomo/KEF (Branch `combo-scripter-v2`) - Get AndroidAPS from https://gitlab.com/jotomo/KEF (Branch `combo-scripter-v2`)
- Make sure you're basal rate is properly set on the pump - Before enabling the Combo plugin in AAPS make sure you're profile is set up
correctly and your basal profile is up to date as AAPS will sync the basal profile
to the pump.
- There might be minor glitches around enabling/disabling the Combo plugi, requiring
to restart AAPS by force-closing it.
Usage: Usage:
- This is not a product, esp. in the beginning the user needs to monitor and understand the system, - This is not a product, esp. in the beginning the user needs to monitor and understand the system,
@ -82,6 +87,9 @@ Usage:
- Read the documentation on the wiki as well as the docs at https://openaps.org - Read the documentation on the wiki as well as the docs at https://openaps.org
- Don't press any buttons on the pump while AAPS communicates with the pump (Bluetooth logo is - Don't press any buttons on the pump while AAPS communicates with the pump (Bluetooth logo is
shown on the pump). shown on the pump).
- If the loop requests a running TBR to be cancelled the Combo will set a TBR of 90% or 110%
for 15 minutes instead. This is because cancelling a TBR causes an alert on the pump which
causes a lot of vibrations.
Reporting bugs: Reporting bugs:
- Note the precise time the problem occurred and describe the circumstances and steps that caused - Note the precise time the problem occurred and describe the circumstances and steps that caused