Improved AAPSClient display of reservoir level and pump status when using OmniPod DASH

This commit is contained in:
Michael Kroes 2022-10-24 16:22:07 +02:00
parent fcd6c68773
commit 1afe060e09
3 changed files with 18 additions and 10 deletions

View file

@ -16,6 +16,7 @@ class DeviceStatusData @Inject constructor() {
var voltage = 0.0
var status = "N/A"
var reservoir = 0.0
var reservoirDisplayOverride = ""
var extended: Spanned? = null
var activeProfileName: String? = null
}

View file

@ -168,16 +168,20 @@ class NSDeviceStatus @Inject constructor(
pumpData.clock + nsSettingsStatus.extendedPumpSettings("urgentClock") * 60 * 1000L < dateUtil.now() -> Levels.URGENT
pumpData.reservoir < nsSettingsStatus.extendedPumpSettings("urgentRes") -> Levels.URGENT
pumpData.isPercent && pumpData.percent < nsSettingsStatus.extendedPumpSettings("urgentBattP") -> Levels.URGENT
!pumpData.isPercent && pumpData.voltage < nsSettingsStatus.extendedPumpSettings("urgentBattV") -> Levels.URGENT
!pumpData.isPercent && pumpData.voltage > 0 && pumpData.voltage < nsSettingsStatus.extendedPumpSettings("urgentBattV") -> Levels.URGENT
pumpData.clock + nsSettingsStatus.extendedPumpSettings("warnClock") * 60 * 1000L < dateUtil.now() -> Levels.WARN
pumpData.reservoir < nsSettingsStatus.extendedPumpSettings("warnRes") -> Levels.WARN
pumpData.isPercent && pumpData.percent < nsSettingsStatus.extendedPumpSettings("warnBattP") -> Levels.WARN
!pumpData.isPercent && pumpData.voltage < nsSettingsStatus.extendedPumpSettings("warnBattV") -> Levels.WARN
!pumpData.isPercent && pumpData.voltage > 0 && pumpData.voltage < nsSettingsStatus.extendedPumpSettings("warnBattV") -> Levels.WARN
else -> Levels.INFO
}
string.append("<span style=\"color:${level.toColor()}\">")
val insulinUnit = rh.gs(R.string.insulin_unit_shortname)
val fields = nsSettingsStatus.pumpExtendedSettingsFields()
if (fields.contains("reservoir")) string.append(pumpData.reservoir.toInt()).append("U ")
if (pumpData.reservoirDisplayOverride != "") {
string.append(pumpData.reservoirDisplayOverride).append("$insulinUnit ")
}
else if (fields.contains("reservoir")) string.append(pumpData.reservoir.toInt()).append("$insulinUnit ")
if (fields.contains("battery") && pumpData.isPercent) string.append(pumpData.percent).append("% ")
if (fields.contains("battery") && !pumpData.isPercent) string.append(Round.roundTo(pumpData.voltage, 0.001)).append(" ")
if (fields.contains("clock")) string.append(dateUtil.minAgo(rh, pumpData.clock)).append(" ")
@ -200,6 +204,7 @@ class NSDeviceStatus @Inject constructor(
deviceStatusPumpData.clock = clock
if (pump.has("status") && pump.getJSONObject("status").has("status")) deviceStatusPumpData.status = pump.getJSONObject("status").getString("status")
if (pump.has("reservoir")) deviceStatusPumpData.reservoir = pump.getDouble("reservoir")
if (pump.has("reservoir_display_override")) deviceStatusPumpData.reservoirDisplayOverride = pump.getString("reservoir_display_override")
if (pump.has("battery") && pump.getJSONObject("battery").has("percent")) {
deviceStatusPumpData.isPercent = true
deviceStatusPumpData.percent = pump.getJSONObject("battery").getInt("percent")

View file

@ -43,13 +43,15 @@ class StatusLightHandler @Inject constructor(
if (pump.pumpDescription.isBatteryReplaceable || pump.isBatteryChangeLoggingEnabled()) {
handleAge(careportal_pb_age, TherapyEvent.Type.PUMP_BATTERY_CHANGE, R.string.key_statuslights_bage_warning, 216.0, R.string.key_statuslights_bage_critical, 240.0)
}
if (!config.NSCLIENT) {
val insulinUnit = rh.gs(R.string.insulin_unit_shortname)
if (pump.model() == PumpType.OMNIPOD_EROS || pump.model() == PumpType.OMNIPOD_DASH) {
handleOmnipodReservoirLevel(careportal_reservoir_level, R.string.key_statuslights_res_critical, 10.0, R.string.key_statuslights_res_warning, 80.0, pump.reservoirLevel, insulinUnit)
} else {
handleLevel(careportal_reservoir_level, R.string.key_statuslights_res_critical, 10.0, R.string.key_statuslights_res_warning, 80.0, pump.reservoirLevel, insulinUnit)
}
if (!config.NSCLIENT) {
if (bgSource.sensorBatteryLevel != -1)
handleLevel(careportal_sensor_battery_level, R.string.key_statuslights_sbat_critical, 5.0, R.string.key_statuslights_sbat_warning, 20.0, bgSource.sensorBatteryLevel.toDouble(), "%")
else
@ -94,7 +96,7 @@ class StatusLightHandler @Inject constructor(
// Omnipod only reports reservoir level when it's 50 units or less, so we display "50+U" for any value > 50
@Suppress("SameParameterValue")
private fun handleOmnipodReservoirLevel(view: TextView?, criticalSetting: Int, criticalDefaultValue: Double, warnSetting: Int, warnDefaultValue: Double, level: Double, units: String) {
if (level > OmnipodConstants.MAX_RESERVOIR_READING) {
if (level >= OmnipodConstants.MAX_RESERVOIR_READING) {
@Suppress("SetTextI18n")
view?.text = " 50+$units"
view?.setTextColor(rh.gac(view.context, R.attr.defaultTextColor))