Update PumpType.java
Ammended the basebasalstep values fro the insight and the combo. this was preventing AAPS from implementing zero basals where the basalstep was higher than the current running basal rate e.g. basal of 0.06u\hr running and a basalstep 0f 0.1u\hr resulted in no zero temp being enacted because of logic in the loop plugin.
This commit is contained in:
parent
21c7da4ae2
commit
167ed27659
|
@ -38,7 +38,7 @@ public enum PumpType {
|
|||
new DoseSettings(0.1d, 15, 12*60, 0.1d), //
|
||||
PumpTempBasalType.Percent,
|
||||
new DoseSettings(10, 15, 12*60,0d, 500d), PumpCapability.BasalRate_Duration15and30minAllowed, //
|
||||
0.01d, 0.1d, DoseStepSize.ComboBasal, PumpCapability.ComboCapabilities), //
|
||||
0.01d, 0.01d, DoseStepSize.ComboBasal, PumpCapability.ComboCapabilities), //
|
||||
|
||||
AccuChekSpirit("Accu-Chek Spirit", 0.1d, null, //
|
||||
new DoseSettings(0.1d, 15, 12*60, 0.1d), //
|
||||
|
@ -50,7 +50,7 @@ public enum PumpType {
|
|||
new DoseSettings(0.05d, 15, 24*60, 0.05d), //
|
||||
PumpTempBasalType.Percent,
|
||||
new DoseSettings(10, 15, 12*60,0d, 250d), PumpCapability.BasalRate_Duration15and30minAllowed, //
|
||||
0.02d, 0.1d, null, PumpCapability.InsightCapabilities), //
|
||||
0.02d, 0.01d, null, PumpCapability.InsightCapabilities), //
|
||||
|
||||
// Animas
|
||||
AnimasVibe("Animas Vibe", 0.05d, null, // AnimasBolus?
|
||||
|
|