2017-11-12 15:59:22 +01:00
|
|
|
- [ ] Bolusing
|
|
|
|
- [ ] Cancelling bolus at various stages and checking error message (if any) and check
|
|
|
|
no bolus or a partial bolus was delivered
|
|
|
|
- [ ] Cancelling a bolus 0.2 U before its finished (cancelling requires pressing up/down
|
|
|
|
3s, so the cancellation will be too late, but the code should handle it and report
|
|
|
|
the full bolus being delivered
|
|
|
|
- [ ] Low cartridge alarm during bolus
|
|
|
|
- [ ] must be confirmed by AAPS
|
|
|
|
- [ ] bolus must have been fully delivered by pump
|
|
|
|
- [ ] bolus must have been added to DB
|
|
|
|
- [ ] the confirmed pump warning must be raised as a notification in AAPS
|
|
|
|
- [ ] notification must also appear on smartwatch
|
|
|
|
- [ ] An error during bolus must yield an error in AAPS
|
|
|
|
- [ ] An error during bolus must yield a notification on a smartwatch
|
|
|
|
- [ ] Bolusing e.g. 4 U if reservoir has only 2 U left must be rejected
|
|
|
|
- [ ] Pressing a button on the pump before bolus delivery started must be handled gracefully
|
|
|
|
- [ ] Same as above, but moving pump out of range
|
|
|
|
- [ ] Pressing a button on the pump after bolus delivery has started must be handled
|
|
|
|
gracefully (bolus finished and AAPS must correctly add it the the DB)
|
|
|
|
- [ ] Same as above, but moving pump out of range
|
|
|
|
- [ ] Giving a bolus on the pump and immediately afterwards entering a bolus on AAPS
|
|
|
|
must reject the bolus with a warning that IOB is not up to date
|
|
|
|
- [ ] A bolus given on the pump must appear in AAPS after the next communication or when
|
|
|
|
manually refreshing via the Combo tab
|
|
|
|
- [ ] Test the highest bolus you'd ever give yourself (AAPS has a configurable limit and the pump
|
|
|
|
has a limit which can be configured with the Config SW)
|
|
|
|
- [ ] BT disconnect issues
|
|
|
|
- [ ] Moving pump out of reach when setting TBR causes "TBR cancelled" alarm on pump.
|
|
|
|
When putting pump close to phone, AAPS must confirm the alert and successfully
|
|
|
|
retry setting TBR
|
|
|
|
- [ ] Same as above while bolusing must report an error and NOT retry the command
|
|
|
|
- [ ] AAPS start
|
|
|
|
- [ ] Starting AAPS without a reachable pump must show something sensible in the Combo tab
|
|
|
|
(that is not hang indefinitely with "initializing" activity)
|
|
|
|
- [ ] Starting AAPS without a reachable pump must trigger "pump unrechable" alert after
|
|
|
|
5 to 10 min.
|
|
|
|
- [ ] If 'sync profile to pump' is set the basal rate profile must be synced to pump on start
|
|
|
|
- [ ] If 'sync profile to pump' is set NOT set, the pump's profile must not be touched
|
|
|
|
- [ ] Read history using Smartpix and compare with AAPS' DB (treatment tab)
|
|
|
|
Esp. those times we communication was interrupted, boluses were cancelled, ...
|
|
|
|
- [ ] Boluses
|
|
|
|
- [ ] TBR
|
|
|
|
- [ ] Alerts
|
|
|
|
- [ ] TDDs
|
|
|
|
- [ ] Date/Time
|
|
|
|
- [ ] Time must be set properly upon app start if pump time is off
|
|
|
|
- [ ] Daylight saving time changes must be detected and pump time updated.
|
|
|
|
Basically any time change greater than a minute on the phone must be picked up
|
|
|
|
by AAPS and result in the pump's time being updated the next time communication
|
|
|
|
with the pump takes place, so disabling automatic time updates on the phone and
|
|
|
|
changing the phone clock manually should allow testing this
|
2017-11-13 22:55:48 +01:00
|
|
|
- [ ] XXX test year change, reading history
|
|
|
|
Set phone time to 31.12, 23:55, press Refresh in Combo tab to sync time to pump,
|
|
|
|
enter bolus
|
|
|
|
- [ ] XXX daylight saving time changes (2am twice)
|
2017-11-12 15:59:22 +01:00
|
|
|
- [ ] Disconnected pump
|
|
|
|
...
|
|
|
|
- [ ] Refilling cartridge
|
|
|
|
- [ ] If TBR was cancelled by refilling, AAPS must detect this and update the TBR treatment
|
|
|
|
accordingly
|
|
|
|
- [ ] Stress testing
|
|
|
|
- PersistentNotification plugin disabled
|
|
|
|
- Lots of comms running, like Wifi, GSM, BT audio
|
|
|
|
- AAPS running in background
|
|
|
|
- Foreground app stresses the phone's memory, CPU (like a game) (potentially) pushing AAPS out of memory
|
|
|
|
- [ ] TBR must still be set/cancelled while running in the background
|
|
|
|
- [ ] With the pump powered off or out of reach, the 'pump unreachable alert' must still
|
|
|
|
trigger
|
|
|
|
- [ ] Combo tab
|
|
|
|
- [ ] Check displayed data (state, battery, reservoir, last bolus, temp basal) is the same
|
|
|
|
as on the pump
|
|
|
|
- [ ] Last connection must indicate issues (yellow if recent connections/commands failed),
|
|
|
|
red if communications failed for 30 or more minutes
|
|
|
|
- [ ] Unsafe usage
|
|
|
|
- [ ] An extended or multiwave bolus given within the last six hour must raise an alert and
|
|
|
|
closed loop functionality (Loop plugin shows "disabled by constrains" when "execute" is pressed)
|
|
|
|
- [ ] Closed loop functionality must resume 6 h after the last ext/multiwave bolus
|
|
|
|
- [ ] An active ext/multiwave bolus (a history record is created only after the bolus completed)
|
|
|
|
must also raise an alert and disable loop
|
|
|
|
- [ ] If a basal rate other than profile 1 is activated, this must also raise an alert and disable
|
|
|
|
the loop for 6 h
|
|
|
|
|